M3000A/M3046AM3015A/M3016A Service Guide
Repairing the Instrument 6-3
6WHS Switch on the Instrument and disconnect the AC power cord from both the Instru-
ment and the mains power source (after 20 seconds, the capacitors on the Power
Supply board are drained).
6WHS Push a thin-bladed screwdriver into the hole next to the cover.
6WHS Release the cover.
6WHS Remove the two anti-vibration screws. (These are the only anti-vibration screws
and must be refitted. Do not lose these screws.)
6WHS Slide out the Power Supply.
5HPRYLQJWKH&KDVVLV
5HDGWKHZDUQLQJVDWWKHEHJLQQLQJRIWKLVFKDSWHU
To remove the chassis, do the following:
6WHS Remove the Server.
6WHS Remove the battery.
6WHS Switch the Instrument on and disconnect the AC power cord from the Instrument
and the mains power source (after 20 seconds, the capacitors on the Power Supply
board are drained).