M3000A/M3046AM3015A/M3016A Service Guide
4-28 Testing the Instrument
63DWLHQW/HDNDJHFXUUHQW6LQJOH)DXOW&RQGLWLRQ6)&PDLQVRQDSSOLHGSDUW
Expected test results:
Maximum leakage current, x = 50µA @ 250V (IEC601-1 and UL2601-1)
Measures patient leakage current from applied Part to earth caused by external main
voltage on applied Part with switch S5 open and closed. Each polarity combination
possible is tested using S2 and S6. This test is applicable for every measurement input (
ECG / RESP, Invasive Pressure, CO
2
, SpO
2
, Temperature)
5HSRUWLQJVDIHW\WHVW6LQWKH6HUYLFHUHFRUG
S(3):P/x
or
S(3):F/x
where P = Pass, F = Fail and x is the value defined in the test described above