Agilent Technologies FS2010 Manual de Serviço Página 192

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 191
184 Agilent Signal Generators Programming Guide
Programming the Status Register System
Status Groups
Baseband Operation Condition Register
The Baseband Operation Condition Register continuously monitors the hardware and firmware status
of the signal generator. Condition registers are read only.
Baseband Operation Transition Filters (negative and positive)
The Baseband Operation Transition Filters specify which types of bit state changes in the condition
register set corresponding bits in the event register. Changes can be positive (0 to 1) or negative (1
to 0).
Table 4-7 Baseband Operation Condition Register Bits
Bit Description
0 Baseband 1 Busy. A 1 in this position indicates the signal generator baseband is active.
1 Baseband 1 Communicating. A 1 in this bit position indicates that the signal generator baseband generator is
handling data IO.
2–14 Unused. This bit position is always set to 0.
15 Always 0.
Query:
STATus:OPERation:BASeband:CONDition?
Response: The decimal sum of the bits set to 1
Example: The decimal value 2 is returned. The decimal sum = 2 (bit 1).
Commands:
STATus:OPERation:BASeband:NTRansition <value> (negative transition), or
STATus:OPERation:BASeband:PTRansition <value> (positive transition), where
<value> is the sum of the decimal values of the bits you want to enable.
Queries:
STATus:OPERation:BASeband:NTRansition?
STATus:OPERation:BASeband:PTRansition?
Vista de página 191
1 2 ... 187 188 189 190 191 192 193 194 195 196 197 ... 395 396

Comentários a estes Manuais

Sem comentários