Chapter 3: Testing Performance
DC Gain Accuracy Test
24
Procedure
1 Make the connections to scope channel 1 as shown below.
Connections
2 Press [Default Setup], then configure the scope as follows:
a Select Setup > Acquisition.
Notes:
Where the BNC Tee adapter is used, it is important to connect it directly to the scope channel input
using the BNC (f) to SMA (m) adapter and the connection saver to minimize ground potential
differences and to ensure that the DMM measures the input voltage to the scope channel as accurately
as possible. Differences in ground potential can be a significant source of measurement error,
particularly at high scope sensitivities.
It also helps to reduce ground potential differences if the scope and DMM are connected to the same
AC supply circuit.
256 averages are used in the scope measurements of this section to reduce measurement noise and to
reduce the measurement error due to resolution.
Oscilloscope Channel Input
Cal Out on Oscilloscope Front Panel
BNC (f) to SMA (m) adapter
BNC (f) to SMA (m) adapter
Connection Saver
Connection Saver