Página 1 - Oscilloscopes

Service GuidePublication Number 54932- 97006October 2013© Agilent Technologies, Inc. 2012- 2013Infiniium 90000 Q- Series Oscilloscopes

Página 2

10CalibrationA calibration is simply an oscilloscope self- adjustment. The purpose of a calibration is performance optimization.There are three ways t

Página 3 - In This Service Guide

Chapter 5: Replacing AssembliesRemoving and Replacing the Acquisition Boards/Backplane Assembly100Removing the Backplane Board from the Deck1 Loosen t

Página 4

Chapter 5: Replacing AssembliesSetting the Calibration Factors after Replacing an Acquisition Board101Setting the Calibration Factors after Replacing

Página 5 - General Information

Chapter 5: Replacing AssembliesRemoving and Replacing the Fans102Removing and Replacing the FansWARNING AVOID INJURY!The fan blades are exposed both i

Página 6

Chapter 5: Replacing AssembliesRemoving and Replacing the Fans1034Clip off the end of each of the fan mounts (see Figure 5- 69) and take the fan out.

Página 7 - Accessories supplied

Chapter 5: Replacing AssembliesRemoving and Replacing the USB Controller Cable104Removing and Replacing the USB Controller Cable1 Remove the top panel

Página 8

Chapter 5: Replacing AssembliesRemoving and Replacing the USB Controller Cable1054Pull the connector cable out of the oscilloscope.Figure 5-725To repl

Página 9

Chapter 5: Replacing AssembliesRemoving and Replacing the USB Controller Cable106

Página 10 - Calibration

6Ordering Replaceable Parts 108Listed Parts 108Unlisted Parts 108Direct Mail Order System 108Exchange Assemblies 108Exploded Views 109Repl

Página 11 - Running a self calibration

108Replaceable Parts This chapter describes how to order replaceable assemblies and parts for the Agilent 90000 Q- Series oscilloscopes. Service suppo

Página 12

Chapter 6: Replaceable PartsExploded Views109Exploded ViewsFront Panel

Página 13 - Testing Performance

Chapter 2: CalibrationRunning a self calibration11Running a self calibrationThe self calibration uses signals generated in the oscilloscope to calibra

Página 14 - Test Equipment

Chapter 6: Replaceable PartsExploded Views110Fan and Acquisition Assembly

Página 15 - •DC Gain Accuracy Test

Chapter 6: Replaceable PartsExploded Views111Oscillator Assembly

Página 16 - Offset Accuracy Test

Chapter 6: Replaceable PartsExploded Views112Power Supply and PC Motherboard

Página 17

Chapter 6: Replaceable PartsExploded Views113External

Página 18 - Record the

Chapter 6: Replaceable PartsReplaceable Parts List114Replaceable Parts ListThe following table is a list of replaceable parts. The information given f

Página 19

Chapter 6: Replaceable PartsReplaceable Parts List115MP16 54932-47405 1 Knob 12 mm slate plainMP17 54932-47410 1 Knob 18 mm slate yellowMP18 54932-474

Página 20 - Offset Gain Test Procedure

Chapter 6: Replaceable PartsReplaceable Parts List116W12 54932-61618 1 Cable - SR 10 GHz backplaneW13 54932-61629 1 Cable - SMA to SMAW14 54932-61628

Página 21 - Channel 1

Chapter 6: Replaceable PartsReplaceable Parts List1178121-1853 1 Power cord - Taiwan - EM-controlled environment required8121-2158 1 Power cord - Camb

Página 22

Chapter 6: Replaceable PartsReplaceable Parts List118

Página 23 - DC Gain Accuracy Test

7Motherboard Assembly 122Acquisition Assembly 123Backplane Assembly 125Oscillator Assembly 129Front Panel Assembly 130RealEdge Technology As

Página 24

Chapter 2: CalibrationRunning a self calibration12Figure 2-1 Calibration Dialog Box 4 Click Start, then follow the instructions on the screen. The ro

Página 25

Chapter 7: Theory of Operation120Figure 7-1 Oscilloscope block diagram

Página 26

121Theory of OperationThis chapter describes the basic structure of the oscilloscope and how the parts interact. The 90000 Q- Series oscilloscopes are

Página 27

Chapter 7: Theory of OperationMotherboard Assembly122Motherboard AssemblyThe motherboard assembly provides all system control and interface functions

Página 28

Chapter 7: Theory of OperationAcquisition Assembly123Acquisition AssemblyThe acquisition assembly consists of four identical acquisition boards. These

Página 29

Chapter 7: Theory of OperationAcquisition Assembly124Figure 7-4 Oak Module and Acquisition ComponentsThe acquisition front end (analog path) starts at

Página 30

Chapter 7: Theory of OperationBackplane Assembly125Backplane AssemblyThe backplane assembly is a device on the PCI Express (PCIe) bus connected by a P

Página 31 - Preset the power meter

Chapter 7: Theory of OperationBackplane Assembly126Main FPGAThe main FPGA (field- programmable gate array) is the only communication link from the osc

Página 32

Chapter 7: Theory of OperationBackplane Assembly127Clock Distribution System Figure 7- 7 shows the timebase system including its circuitry.Figure 7-7

Página 33

Chapter 7: Theory of OperationBackplane Assembly128The fundamental sample clock is a single- ended 10 GHz signal created by the timebase generation sy

Página 34

Chapter 7: Theory of OperationOscillator Assembly129Oscillator AssemblyThe oscillator assembly consists of an oscillator board; one or two DROs (diele

Página 35

3Vertical Performance Verification 15Offset Accuracy Test 16DC Gain Accuracy Test 23Analog Bandwidth - - Maximum Frequency Test 28Time Scale

Página 36

Chapter 7: Theory of OperationFront Panel Assembly130Front Panel AssemblyThe front panel assembly consists of an LCD display board with backlight; a t

Página 37 - Equipment Required

Chapter 7: Theory of OperationFront Panel Assembly131

Página 38

Chapter 7: Theory of OperationRealEdge Technology Assembly132RealEdge Technology AssemblyThe RealEdge Technology assembly is included with Q- Series o

Página 39 - Agilent Technologies

Chapter 7: Theory of OperationPower Supply133Power SupplyThe power supply is a +12 V bulk supply that is removable, exchangeable, and safety-approved.

Página 40

Chapter 7: Theory of OperationPower Supply134

Página 41

Safety NoticesThis apparatus has been designed and tested in accor-dance with IEC Publication EN 61010-1:2001, Safety Requirements for Measuring Appar

Página 42

Notices© Agilent Technologies, Inc. 2013No part of this manual may be reproduced in any form or by any means (including electronic storage and retriev

Página 43

14This section describes performance test procedures. Performance Test IntervalThe procedures in this section may be performed for incoming inspection

Página 44

Chapter 3: Testing Performance15 Vertical Performance VerificationThis section contains the following vertical performance verification tests:• Offset

Página 45 - Troubleshooting

16Offset Accuracy TestCAUTION Ensure that the input voltage to the oscilloscope never exceeds ±5 V.SpecificationsEquipment RequiredNOTE: The offset ac

Página 46

Chapter 3: Testing PerformanceOffset Accuracy Test17b When the Acquisition Setup window appears, enable averaging and set the # of averages to 256 as

Página 47 - Primary Trouble Isolation

Chapter 3: Testing PerformanceOffset Accuracy Test18c When the Enter Measurement Info window appears, ensure that the V avg function is set up as foll

Página 48 - Check the display

Chapter 3: Testing PerformanceOffset Accuracy Test196 Repeat step 5 for the remaining vertical sensitivities for channel 1 in the Zero Error Test sect

Página 49

2The Agilent Technologies Infiniium 90000 Oscilloscope at a GlanceEase of use with high performanceThe Agilent Technologies Infiniium oscilloscopes co

Página 50

Chapter 3: Testing PerformanceOffset Accuracy Test20Offset Gain Test Procedure1 Make the connections to scope channel 1 as shown below.Connections 2 S

Página 51 - Run a self calibration

Chapter 3: Testing PerformanceOffset Accuracy Test21is set up as follows and then click OK:Source = Channel 1Measurement area = Entire Display4Set the

Página 52 - Power Supply Troubleshooting

Chapter 3: Testing PerformanceOffset Accuracy Test227 Change the channel 1 offset value to - 400.0 mV.8 Set the Cal Out voltage to - 400.0 mV.9 Press

Página 53 - Chapter 4: Troubleshooting

Chapter 3: Testing PerformanceDC Gain Accuracy Test23DC Gain Accuracy TestCAUTION Ensure that the input voltage to the oscilloscope never exceeds ±5 V

Página 54

Chapter 3: Testing PerformanceDC Gain Accuracy Test24Procedure1 Make the connections to scope channel 1 as shown below.Connections 2 Press [Default S

Página 55

Chapter 3: Testing PerformanceDC Gain Accuracy Test25b When the Acquisition Setup window appears, enable averaging and set the # of averages to 256 as

Página 56 - Setting Up the BIOS

Chapter 3: Testing PerformanceDC Gain Accuracy Test26c Select the Vavg measurement as shown below. d When the Enter Measurement Info window appears, e

Página 57 - Motherboard Verification

Chapter 3: Testing PerformanceDC Gain Accuracy Test276 Change the Cal Out voltage to - 30 mV.7 Press [Clear Display] on the scope, wait for the #Avgs

Página 58 - Display Troubleshooting

Chapter 3: Testing PerformanceAnalog Bandwidth -- Maximum Frequency Test28Analog Bandwidth - - Maximum Frequency TestCAUTION Ensure that the input vo

Página 59

Chapter 3: Testing PerformanceAnalog Bandwidth -- Maximum Frequency Test29Connector Saver 3.5 mm (f)-(f), shipped with 20, 25, and 32 GHz oscilloscope

Página 60

3In This Service GuideThis guide provides the service information for the Agilent Technologies 90000 Q- Series oscilloscopes. It is divided into seven

Página 61

Chapter 3: Testing PerformanceAnalog Bandwidth -- Maximum Frequency Test30Connections for all oscilloscope models, channels 1- 4Notes• Connect output

Página 62 - Keyboard Troubleshooting

Chapter 3: Testing PerformanceAnalog Bandwidth -- Maximum Frequency Test31Connections for 50 and 62 GHz oscilloscope models, channels 1R- 3RProcedure1

Página 63 - LED Troubleshooting

Chapter 3: Testing PerformanceAnalog Bandwidth -- Maximum Frequency Test32d Select Setup > Acquisition and then set up the acquisition parameters a

Página 64 - Touch Screen Troubleshooting

Chapter 3: Testing PerformanceAnalog Bandwidth -- Maximum Frequency Test33f When the Enter Measurement Info window appears, configure this measurement

Página 65

Chapter 3: Testing PerformanceAnalog Bandwidth -- Maximum Frequency Test348 Measure the input power to the scope channel and convert this measurement

Página 66 - Checking Probe Power Outputs

Chapter 3: Testing PerformanceAnalog Bandwidth -- Maximum Frequency Test3511 Change the generator frequency to the maximum value for the model being t

Página 67 - Before You Contact Agilent

Chapter 3: Testing PerformanceAnalog Bandwidth -- Maximum Frequency Test3616 Change the scope setup as follows:a Change the channel vertical sensitivi

Página 68

Chapter 3: Testing PerformanceTime Scale Accuracy (TSA) Test37Time Scale Accuracy (TSA) TestThis procedure verifies the maximum TSA specification for

Página 69 - Replacing Assemblies

Chapter 3: Testing PerformanceTime Scale Accuracy (TSA) Test38ConnectionsConnect the equipment as shown in the following figure.Procedure1 Configure t

Página 70

Chapter 3: Testing PerformancePerformance Test Record39Performance Test RecordOffset Accuracy TestZero Error TestAgilent TechnologiesAgilent 90000 Q-S

Página 72

Chapter 3: Testing PerformancePerformance Test Record40Offset Gain TestVertical SensitivityVCal OutSettingVDMM+VDMM-VDMM0VScope+VScope-VScope0Calc. Of

Página 73

Chapter 3: Testing PerformancePerformance Test Record41DC Gain Accuracy TestVertical SensitivityVCal OutSettingVDMM+VDMM-VScope+VScope-Calc. DC Gain E

Página 74

Chapter 3: Testing PerformancePerformance Test Record42Analog Bandwidth - Maximum Frequency CheckMax frequency: DSOX/DSAX92004Q = 20 GHz, DSOX/DSAX925

Página 75

Chapter 3: Testing PerformancePerformance Test Record43Time Scale AccuracyMeasured Time Scale Error (ppm)Years Since Calibration (years)Low Test Limit

Página 76

Chapter 3: Testing PerformancePerformance Test Record44

Página 77 - Pull the hard drive out

4Primary Trouble Isolation 47Power Supply Troubleshooting 52Setting Up the BIOS 56Motherboard Verification 57Display Troubleshooting 58Acqui

Página 78

46TroubleshootingThe service strategy for troubleshooting oscilloscopes is to isolate problems to a faulty assembly, then use the disassembly and asse

Página 79

Chapter 4: TroubleshootingPrimary Trouble Isolation47Primary Trouble IsolationThe main procedural tool in this chapter is the flowchart. The Primary T

Página 80 - Remove the bottom panel

Chapter 4: TroubleshootingPrimary Trouble Isolation48APerform power-up.Connect the oscilloscope power cord and press the power button in the lower lef

Página 81

Chapter 4: TroubleshootingPrimary Trouble Isolation49DCheck the front panel response by running the Keyboard and LED self tests.Follow these steps to

Página 82 - Removing a RealEdge Assembly

1Instruments covered by this service guide 6Accessories supplied 7Specifications and Characteristics 7General Information

Página 83 - Remove the four M3 screws

Chapter 4: TroubleshootingPrimary Trouble Isolation50The Front Panel LED Test screen appears, showing a symbolic representation of all front panel LED

Página 84

Chapter 4: TroubleshootingPrimary Trouble Isolation51ERun a self calibration.1 Complete a self calibration by following the procedures in chapter 2.2

Página 85

Chapter 4: TroubleshootingPower Supply Troubleshooting52Power Supply Troubleshooting This section provides information to help you isolate the problem

Página 86

Chapter 4: TroubleshootingPower Supply Troubleshooting53 LED Type of Fault What To DoPS01 PS02 PS03 PS04ON ON ON ON Supply lateReplace Backplane Assem

Página 87

Chapter 4: TroubleshootingPower Supply Troubleshooting54The second set of LEDs is located to the left of the first set and slightly lower, below the J

Página 88

Chapter 4: TroubleshootingPower Supply Troubleshooting55LED Type of Fault What To DoAOT0* Over-temperature Replace acquisition assembly 4AOT1* Over-te

Página 89

Chapter 4: TroubleshootingSetting Up the BIOS56Setting Up the BIOSIf the BIOS settings become corrupt, the Infiniium oscilloscope PC motherboard will

Página 90

Chapter 4: TroubleshootingMotherboard Verification57Motherboard VerificationIf you have been through the Power Supply Troubleshooting section of this

Página 91

Chapter 4: TroubleshootingDisplay Troubleshooting58 Display TroubleshootingFigure 4-8 Display Troubleshooting Flowchart&RQQHFWH[WHUQDOPRQLWRU

Página 92

Chapter 4: TroubleshootingDisplay Troubleshooting59WARNING INJURY CAN RESULT!Use caution when the oscilloscope fan blades are exposed as they can caus

Página 93

6Instruments covered by this service guideOscilloscopes manufactured after the date this manual was released may be different from those described in

Página 94 - Assembly

Chapter 4: TroubleshootingAcquisition/Backplane Assembly Troubleshooting60Acquisition/Backplane Assembly TroubleshootingThis section describes which b

Página 95

Chapter 4: TroubleshootingAcquisition/Backplane Assembly Troubleshooting61Merwig6, Merwig7, Hedwig12, Hedwig13, Hedwig14, Hedwig15, Oak6Replace acquis

Página 96

Chapter 4: TroubleshootingKeyboard Troubleshooting62 Keyboard TroubleshootingIf some of the knobs fail the keyboard self test but some work properly,

Página 97 - Removing an Acquisition Board

Chapter 4: TroubleshootingLED Troubleshooting63LED TroubleshootingLED Failure During System Start- UpWhen power is turned on to the scope, the front p

Página 98

Chapter 4: TroubleshootingTouch Screen Troubleshooting64Touch Screen TroubleshootingFigure 4-9 Touch screen flowchart A To find the driver, bring up t

Página 99

Chapter 4: TroubleshootingOscillator Board Troubleshooting65Oscillator Board TroubleshootingIf you encounter any problems with the oscillator board, f

Página 100

Chapter 4: TroubleshootingChecking Probe Power Outputs66Checking Probe Power OutputsProbe power outputs are on the front panel, in the lower left corn

Página 101

Chapter 4: TroubleshootingBefore You Contact Agilent67Before You Contact AgilentIf you have read this Troubleshooting chapter and have unresolved ques

Página 102

Chapter 4: TroubleshootingBefore You Contact Agilent68

Página 103

5ESD Precautions 70Tools Required 70Returning the Oscilloscope to Agilent Technologies for Service 71Removing and Replacing the Top Panel and Mo

Página 104 - Remove the top panel

Chapter 1: General InformationAccessories supplied7Accessories suppliedThe following accessories are supplied.•Mouse•Stylus•Keyboard• Front panel cove

Página 105

70Replacing AssembliesUse the procedures in this chapter when removing and replacing assemblies and parts in the Agilent Technologies oscilloscopes.In

Página 106

Chapter 5: Replacing AssembliesReturning the Oscilloscope to Agilent Technologies for Service71Returning the Oscilloscope to Agilent Technologies for

Página 107 - Replaceable Parts

Chapter 5: Replacing AssembliesRemoving and Replacing the Top Panel and Motherboard72Removing and Replacing the Top Panel and MotherboardNote that all

Página 108

Chapter 5: Replacing AssembliesRemoving and Replacing the Top Panel and Motherboard733Disconnect the PCIe cable by pressing down on the locking tab an

Página 109 - Exploded Views

Chapter 5: Replacing AssembliesRemoving and Replacing the Top Panel and Motherboard746Disconnect the two USB cables.Figure 5-77Disconnect the power ca

Página 110 - Chapter 6: Replaceable Parts

Chapter 5: Replacing AssembliesRemoving and Replacing the Top Panel and Motherboard75Figure 5-108If you need to replace the motherboard fan, disconnec

Página 111

Chapter 5: Replacing AssembliesRemoving and Replacing the Top Panel and Motherboard7610Remove the PCIe cable from the motherboard and set it aside to

Página 112

Chapter 5: Replacing AssembliesRemoving and Replacing the Hard Disk Drive77Removing and Replacing the Hard Disk Drive1 Loosen the two screws on back o

Página 113

Chapter 5: Replacing AssembliesRemoving and Replacing the Power Supply78Removing and Replacing the Power SupplyWARNING SHOCK HAZARD! If the power supp

Página 114 - Replaceable Parts List

Chapter 5: Replacing AssembliesRemoving and Replacing the Front Panel Assembly and Bottom Panel79Removing and Replacing the Front Panel Assembly and B

Página 115

Chapter 1: General InformationSpecifications and Characteristics8

Página 116

Chapter 5: Replacing AssembliesRemoving and Replacing the Front Panel Assembly and Bottom Panel803Turn the oscilloscope upside- down and remove the fo

Página 117

Chapter 5: Replacing AssembliesRemoving and Replacing the Front Panel Assembly and Bottom Panel816Disconnect each input cable from each of the four at

Página 118

Chapter 5: Replacing AssembliesRemoving and Replacing the Front Panel Assembly and Bottom Panel829Disconnect the Aux Out and Cal Out cables from both

Página 119 - Theory of Operation

Chapter 5: Replacing AssembliesRemoving and Replacing the Front Panel Assembly and Bottom Panel832Disconnect the cable connecting the pilot cables to

Página 120

Chapter 5: Replacing AssembliesRemoving and Replacing the Front Panel Assembly and Bottom Panel845Gently remove the RealEdge assembly, keeping the cab

Página 121

Chapter 5: Replacing AssembliesRemoving and Replacing the Front Panel Assembly and Bottom Panel85Finishing Removal of the Front Panel1 To finish remov

Página 122 - Motherboard Assembly

Chapter 5: Replacing AssembliesRemoving and Replacing Front Panel Assembly Parts86Removing and Replacing Front Panel Assembly PartsRemove the front pa

Página 123 - Acquisition Assembly

Chapter 5: Replacing AssembliesRemoving and Replacing Front Panel Assembly Parts873Remove the six screws from the back of the front panel.Figure 5-364

Página 124

Chapter 5: Replacing AssembliesRemoving and Replacing Front Panel Assembly Parts886The front circuit board of the keyboard assembly is connected to th

Página 125 - Backplane Assembly

Chapter 5: Replacing AssembliesRemoving and Replacing Front Panel Assembly Parts89Cal Connector (Power Button) Assembly1 Remove the two screws connect

Página 126 - Main FPGA

2Running a self calibration 11Calibration

Página 127 - 7LPHEDVH

Chapter 5: Replacing AssembliesRemoving and Replacing Front Panel Assembly Parts903Remove the four M3 screws (two on each side) connecting the display

Página 128 - Trigger System

Chapter 5: Replacing AssembliesRemoving and Replacing RealEdge Components91Removing and Replacing RealEdge ComponentsThis section describes how to rem

Página 129 - 2VFLOODWRU%RDUG

Chapter 5: Replacing AssembliesRemoving and Replacing RealEdge Components924Carefully remove the RealEdge module. Figure 5-485Remove the two screws on

Página 130 - Front Panel Assembly

Chapter 5: Replacing AssembliesRemoving and Replacing RealEdge Components938Remove the two screws connecting the attenuator to the bracket. When repla

Página 131

Chapter 5: Replacing AssembliesRemoving and Replacing the Acquisition Boards/Backplane Assembly94Removing and Replacing the Acquisition Boards/Backpla

Página 132 - RealEdge Technology Assembly

Chapter 5: Replacing AssembliesRemoving and Replacing the Acquisition Boards/Backplane Assembly955Disconnect the cables from the rear panel (four cabl

Página 133 - Power Supply

Chapter 5: Replacing AssembliesRemoving and Replacing the Acquisition Boards/Backplane Assembly968Remove the four M4 screws from the rear panel of the

Página 134

Chapter 5: Replacing AssembliesRemoving and Replacing the Acquisition Boards/Backplane Assembly97Removing an Acquisition BoardWhen removing an individ

Página 135 - Safety Notices

Chapter 5: Replacing AssembliesRemoving and Replacing the Acquisition Boards/Backplane Assembly983Remove the 10 M4 screws (five from each side) connec

Página 136

Chapter 5: Replacing AssembliesRemoving and Replacing the Acquisition Boards/Backplane Assembly997Remove the eight remaining screws securing the acqui