6
Bluetooth EDR receiver tests
EDR Rx sensitivity
Link conditions
Link mode Test mode (loopback)
Payload PRBS9, BSOO, BSFF, BS55
Packet type 2-DH1, 2-DH3, 2-DH5, 3-DH1,
3-DH3, 3-DH5
Number of bits 1 to 200,000,000
Impairments
Frequency offset ±100 kHz
Frequency offset resolution 1 kHz
Symbol timing –30 to +30 ppm
Symbol timing resolution 1 ppm
“Dirty transmitter” impairments for EDR Rx sensitivity
measurements
Set of
parameters
Carrier offset
frequency (kHz)
Symbol timing
offset (ppm)
1 0 0
2 +65 +20
3 –65 –20
Sine impairments for EDR Rx sensitivity measurements
Modulation frequency range 300 Hz to 10 kHz
Resolution 100 Hz
Maximum deviation range 0 Hz to 40 kHz
Resolution 1 kHz
Measurement
Supported measurements BER, number of bit errors, number
of Rx bits, PER, number of NACK
packets, number of errored packets,
number of Tx packets, number of
HEC, CRC, and NACK error
Number of measurement
channels
1
3, hopping
Range 0 to –90 dBm
Resolution 0.1 dB
Accuracy
2, 3
±0.6 dB, –35 to –90 dBm
±1 dB, > 35 dBm
EDR Rx BER floor sensitivity
Link conditions
Link mode Test mode (loopback)
Hopping On or off
Payload PRBS9, BSOO, BSFF, BS55
Packet type 2-DH1, 2-DH3, 2-DH5, 3-DH1,
3-DH3, 3-DH5
Number of bits 1 to 200,000,000
Measurement
Supported measurements BER, number of bit errors, number
of Rx bits, PER, number of NACK
packets, number of errored packets,
number of Tx packets, number of
HEC, CRC, and NACK error
Number of measurement
channels
1
3, hopping
Range 0 to –90 dBm
Resolution 0.1 dB
Accuracy
2, 3
±0.6 dB, –35 to –90 dBm
±1 dB, > 35 dBm
EDR Rx maximum input level
Link conditions
Link mode Test mode (loopback)
Hopping On or off
Payload PRBS9, BSOO, BSFF, BS55
Packet type 2-DH1, 2-DH3, 2-DH5, 3-DH1,
3-DH3, 3-DH5
Number of bits 1 to 200,000,000
Measurement
Supported measurements BER, number of bit errors, number
of Rx bits, PER, number of NACK
packets, number of errored packets,
number of Tx packets, number of
HEC, CRC, and NACK error
Number of measurement
channels
1
3, hopping
Range 0 to –90 dBm
Resolution 0.1 dB
Accuracy
2, 3
±0.6 dB, –35 to –90 dBm
±1 dB, > 35 dBm
1. Internal sequencer enables three measurement channels to be measured
consecutively. Measurements on all 79 Bluetooth channels are supported.
2. Verified using CW measurements.
3. Add 0.01 dB/°C from 30 to 55 °C, add 0.07 dB/°C from 20 to 0 °C.