Getting Started Guide Wireless Connectivity Test SetN4010AManufacturing Part Number: N4010-90086Printed in MalaysiaApril 2014© Copyright Agilent Techn
Getting StartedWelcomeChapter 12WelcomeWelcome to the Agilent Technologies N4010A Wireless Connectivity Test Set’s Getting Started Guide.Use this guid
Getting StartedInterfacing with the N4010AChapter 1 3Interfacing with the N4010ASome of the measurments require that a Dynamic Link Library (DLL) is i
Getting StartedInterfacing with the N4010AChapter 14Figure 1-1 N4010A Interfaces and DLLsControl of Test Devicevia LMP commands from Test Set and EUT
Getting StartedInitial InspectionChapter 1 5Initial InspectionPlease inspect the shipping container for damage. If the shipping container or packaging
Getting StartedAdjusting the Carrying HandleChapter 16Adjusting the Carrying Handle Adjust the carrying handle to carry the Test Set or view the displ
Getting StartedTurning the Test Set OnChapter 1 7Turning the Test Set OnCAUTION The Test Set has an autoranging power supply. Ensure the supply voltag
Getting StartedTurning the Test Set OnChapter 18NOTE If the test set has been stored in extremely cold conditions, beyond its normal operating range,
Getting StartedTurning the Test Set OnChapter 1 9System InformationTo verify your configuration, press the System key, an example is shown in Figure 1
Getting StartedTurning the Test Set OnChapter 110What Can Go Wrong?See this Do thisRed LED not lit Check that power is supplied to the test set.Check
Getting StartedMinimum PC RequirementsChapter 1 11Minimum PC RequirementsSome N4010A TestSet methods of operation require a higher preformance PC than
iiGeneralThe material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to
Getting StartedInstalling the Measurement Software, Applications, and HelpChapter 112Installing the Measurement Software, Applications, and HelpThe su
Chapter 2 132 Making Your First MeasurementsN4010-90086.book Page 13 Friday, April 4, 2014 3:12 PM
Making Your First MeasurementsSelecting the Application ModeChapter 214Selecting the Application ModeThe purpose of this guide is to allow you to conf
Making Your First MeasurementsRunning a Confidence Check on the Front Panel (Bluetooth)Chapter 2 15Running a Confidence Check on the Front Panel (Blue
Making Your First MeasurementsRunning a Confidence Check on the Front Panel (Bluetooth)Chapter 216Figure 2-2 Test Set RFIO Connections3. Press the Con
Making Your First MeasurementsRunning a Confidence Check on the Front Panel (Bluetooth)Chapter 2 17Figure 2-4 Typical Inquiry Results Screen8. Ensure
Making Your First MeasurementsConnecting USB, GPIB or LANChapter 218Connecting USB, GPIB or LANThe Test Set can be remotely controlled by USB, GPIB (I
Making Your First MeasurementsConnecting USB, GPIB or LANChapter 2 19Connecting by USBThe USB interface requires no front panel configuration. USB ope
Making Your First MeasurementsConnecting USB, GPIB or LANChapter 220Connecting by GPIB Changing the GPIB AddressThe GPIB address is an integer between
Making Your First MeasurementsConnecting USB, GPIB or LANChapter 2 21Figure 2-6 Typical Connection Expert Screen showing GPIB Connection4. The Test Se
iiiLegal InformationCertificationAgilent Technologies certifies that this product met its published specifications at the time of shipment from the
Making Your First MeasurementsConnecting USB, GPIB or LANChapter 222Connecting by LAN The Test Set has two LAN operating modes.• Dynamic Host Configu
Making Your First MeasurementsConnecting USB, GPIB or LANChapter 2 23DHCP ModeIn DHCP mode the IP Address, Subnet Mask and Default Gateway values are
Making Your First MeasurementsConnecting USB, GPIB or LANChapter 224Configuring the Connection Expert1. Run the Connection Expert program. It can be o
Making Your First MeasurementsConnecting USB, GPIB or LANChapter 2 256. In the pop up screen, shown in Figure 2-9, select the IP Address link.7. Enter
Making Your First MeasurementsConnecting USB, GPIB or LANChapter 226Static IPIn static mode you must set up the IP Address, Subnet Mask, and Default G
Making Your First MeasurementsRunning a SCPI Confidence CheckChapter 2 27Running a SCPI Confidence CheckIf you have previously installed the Agilent I
Making Your First MeasurementsRunning a SCPI Confidence CheckChapter 228Step 3. Sending SCPI commands1. In the Agilent Connection Expert program, clic
Making Your First MeasurementsRunning a WLAN Confidence CheckChapter 2 29Running a WLAN Confidence CheckTo operate a WLAN Confidence Check, your N4010
Making Your First MeasurementsRunning a WLAN Confidence CheckChapter 230If you have previously installed the N4010A Wireless LAN Test Suite, the Agile
Making Your First MeasurementsRunning a WLAN Confidence CheckChapter 2 31Step 3. Rename the Test Set in Connection Wizard.1. In the Agilent Connection
ivSafety NoticesThis guide uses warnings and cautions to denote hazardsWARNING Warning denotes a hazard. It calls attention to a procedure, which if n
Making Your First MeasurementsRunning a WLAN Confidence CheckChapter 2324. Click on the Demo tab.5. Click on Run Demo.6. The display shows a loopback
Making Your First MeasurementsRunning the Graphical Measurement Application (GMA) N4017A.Chapter 2 33Running the Graphical Measurement Application (GM
Making Your First MeasurementsRunning the Graphical Measurement Application (GMA) N4017A.Chapter 234If you have previously installed the Agilent N4017
Making Your First MeasurementsRunning the Graphical Measurement Application (GMA) N4017A.Chapter 2 35Step 3. Connecting the Test SetRefer to “Connecti
Making Your First MeasurementsRunning the Graphical Measurement Application (GMA) N4017A.Chapter 2365. Use the Instrument Selector to locate your Test
Making Your First MeasurementsRunning the Graphical Measurement Application (GMA) N4017A.Chapter 2 37Figure 2-18 Typical Connection Control Screen10.
Making Your First MeasurementsRunning the Graphical Measurement Application (GMA) N4017A.Chapter 23815. In the Control menu, click Run Single.16. The
Making Your First MeasurementsRack Mounting the Test SetChapter 2 39Rack Mounting the Test SetDisconnect Device RequirementsIn rack mounting operation
Making Your First MeasurementsRack Mounting the Test SetChapter 2403. Fit the Rack Mount Flanges.(Parts available separately: 5063-9241).Ready for ins
Chapter 3 413 Regulatory InformationN4010-90086.book Page 41 Friday, April 4, 2014 3:12 PM
vflammable gasses or fumes.DO NOT perform procedures involving cover or shield removal unless you are qualified to do so: Operating personnel must n
Regulatory InformationGeneral SpecificationsChapter 342General SpecificationsEnvironmentalPhysical SpecificationsPower RequirementsCooling Requirement
Regulatory InformationGeneral SpecificationsChapter 3 43CleaningUse a soft, clean, damp cloth to clean the front-panel and side covers.UsageThis Test
Regulatory InformationCompliance and MarkingsChapter 344Compliance and MarkingsElectromagnetic Compatibility (EMC)This product conforms with the prote
Regulatory InformationCompliance and MarkingsChapter 3 45The CSA mark is a registered trademark of the Canadian Standards Association and indicates co
Regulatory InformationRegulatory InformationChapter 346Regulatory InformationSound EmissionHerstellerbescheinigungDiese Information steht im Zusammenh
Regulatory InformationResponsibilities of the CustomerChapter 3 47Responsibilities of the CustomerThe customer shall provide:• Access to the products
Regulatory InformationAgilent Sales and Service OfficesChapter 348Agilent Sales and Service OfficesIn any correspondence or telephone conversations, r
viWARNING Appliance coupler (mains input power cord) is the power disconnect device. Do not position the Test Set such that access to the coupler is i
Contents vii1. Getting StartedWelcome . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ContentsviiiN4010-90086.book Page viii Friday, April 4, 2014 3:12 PM
Chapter 1 11 Getting StartedN4010-90086.book Page 1 Friday, April 4, 2014 3:12 PM