-440 Query UNTERMINATED after indefinite response
A query was received in the same program message in which a query
requesting an indefinite response was executed (see IEEE 488.2,
6.3.7.5).
-430 Query DEADLOCKED
A SCPI output queue has filled, preventing further SCPI command
execution, and there is no more room left in the corresponding SCPI
input queue to accept a query to read from the output queue. The
system automatically discards output to correct the deadlock.
-420 Query UNTERMINATED
A condition causing an UNTERMINATED query error occurred (see
IEEE 488.2, 6.3.2.2). For example, the device was addressed to
talk and an incomplete program message was received.
-410 Query INTERRUPTED
A condition causing an INTERRUPTED query error occurred (see
IEEE 488.2, 6.3.2.7). For example, a query was followed by DAB
or GET before a response was completely sent.
-400 Query Error
This is a generic query error for devices that cannot detect
more specific errors. The code indicates only that a query
error as defined in IEE 488.2, 11.5.1.1.7 and 6.3 has occurred.
-350 Queue overflow
This is a specific code entered into the queue in lieu of the code
that caused the error. This message indicates that there is no more
room in the queue and an error occurred but was not recorded.
-330 Self-test failed; BBG{value}: {name}
A self-test executed on the baseband generator failed. Report this
error to Agilent Technologies.
-330 Self-test failed; The demodulator board failed to power up properly.
The demodulator board failed to power up properly. Report this
error to the nearest Agilent Technologies sales and service office.
-330 Self-test failed; The loop back test failed because the digital module
could not be set up.
To run a diagnostic loop back test, the instrument needs to
communicate with and set up the module. This error is reported when
the instrument cannot communicate with the module.
-321 Out of memory
*** No documentation ***
-321 Out of memory; Arb file selection failed. File System failed to access
required file. Some table editor files may be corrupted.
There was a problem accessing a required file.