Getting Started
Configuring for Remote Control
Chapter 2 13
6. Press
Enter > Proceed With Reconfiguration > Confirm Change (Instrument will Reboot).
The signal generator will reboot with the revised settings saved.
This assigns a hostname and IP address to the signal generator. The hostname and IP address
are persistent states; they are not affected by an instrument preset or a power cycle.
RS-232 Interface Configuration
1. Press Utility > GPIB/RS-232 LAN > RS-232 Setup.
2. Press
RS-232 Baud Rate.
3. Press the desired baud rate softkey.
4. Press
RS-232 Echo Off On.
This toggles the state of the SCPI echoing on the RS-232 connection. Set as desired.
5. Press
Reset RS-232.
This deletes the data from the RS-232 buffer, discarding any unprocessed SCPI input
received over RS-232.
The RS-232 parameters are persistent states; these states are not affected by an instrument
preset or power cycle.