Chapter 3 53
Programming Command Reference
CONFigure Subsystem
3. Programming Command
Reference
RX Level Test
CONFigure:SEQuence:LEVel <int>
CONFigure:SEQuence:LEVel?
This command defines whether or not to include RX Level tests in the test
sequence for the six measurements. <int> is a binary coded decimal ranging from 0
to 63 as follows:
bit 0 (1): For testing TCH1 or DCP1
bit 1 (2): For testing TCH2 or DCP2
bit 2 (4): For testing TCH3 or DCP3
bit 3 (8): For testing TCH4 or DCP4
bit 4 (16): For testing TCH5 or DCP5
bit 5 (32): For testing TCH6 or DCP6
The choices for each bit are 1 for Run and 0 for --- (skip). At preset, this is set to
63.
RX Quality Test
CONFigure:SEQuence:QUALity <int>
CONFigure:SEQuence:QUALity?
This command defines whether or not to include RX Quality tests in the test
sequence for the six measurements. <int> is a binary coded decimal ranging from 0
to 63 as follows:
bit 0 (1): For testing TCH1 or DCP1
bit 1 (2): For testing TCH2 or DCP2
bit 2 (4): For testing TCH3 or DCP3
bit 3 (8): For testing TCH4 or DCP4
bit 4 (16): For testing TCH5 or DCP5
bit 5 (32): For testing TCH6 or DCP6
The choices for each bit are 1 for Run and 0 for --- (skip). At preset, this is set to
63.
Sensitivity Test
CONFigure:SEQuence:SENSitivity <int>
CONFigure:SEQuence:SENSitivity?
This command defines whether or not to include Sensitivity tests in the test
sequence for the six measurements. <int> is a binary coded decimal ranging from 0
to 63 as follows:
bit 0 (1): For testing TCH1 or DCP1