12 Chapter2
Setting up the HP/Agilent 8922 GSM HSCSD Test System
Enabling the HSCSD functionality
Figure 2-4 Tests screen
Step 4. Set Location to CARD.
Step 5. Select the Procedure field and select ENABLE as the procedure.
Step 6. Select the Run Test field or press the key on the front panel of the
HP/Agilent 8922 to run the program.
The option HSCSD Master is now present in the Instrument Type field
in the CONFIGURE screen indicating that the HSCSD functionality is
correctly enabled.