Agilent Technologies 81640A Guia do Utilizador Página 84

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 114
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 83
Performance Tests Test Recor d
84 Agilent 81689A, 81689B and 81649A Tunable Laser Modules Users Guide, Third Edition
Agilent 81649A Performance Test Page 4 of 8
Model Agilent 81649A Tunable Laser Report No. ________ Date_______
Relative Wavelength Accuracy Largest Maximum Deviation ________nm
Summary of All Repetitions Smallest Minimum Deviation _______nm
Relative Wavelength Accuracy Result (= Largest Maximum Deviation Smallest Minimum Deviation)
Relative Wavelength Accuracy _______nm
Specification: 0.3 nm
Measurement Uncertainty: ± 0.2
pm
Absolute Wavelength Accuracy
Result
Largest Value of Deviation (= largest value of either Largest Maximum
Deviation or Smallest Minimum Deviation)
Absolute Wavelength Accuracy _______nm
Test Limit: 1.0 nm
Specification: 0.6 nm typical
Measurement Uncertainty: ± 0.6 pm
Wavelength Repeatability
Measurement Uncertainty: ± 0.1 pm
Repeatability of 1570.000 nm
(= reference)
Measurement
Result
Repeatability of 1595.000 nm
(= reference)
Measurement
Result
Initial Setting REF = nm Initial Setting REF = nm
from 1580.000 nm to REF nm from 1570.000 nm to REF nm
from 1590.000 nm to REF nm from 1580.000 nm to REF nm
from 1600.000 nm to REF nm from 1590.000 nm to REF nm
from 1610.000 nm to REF nm from 1610.000 nm to REF nm
from 1620.000 nm to REF nm from 1620.000 nm to REF nm
largest measured wavelength nm largest measured wavelength nm
smallest measured wavelength nm smallest measured wavelength nm
Wavelength Repeatability nm Wavelength Repeatability nm
= largest measured wavelength - smallest measured wavelength = largest measured wavelength - smallest measured wavelength
Specification 0.10 nm Specification 0.10 nm
Vista de página 83
1 2 ... 79 80 81 82 83 84 85 86 87 88 89 ... 113 114

Comentários a estes Manuais

Sem comentários