Performance Tests Performance Test Instructions
50 Agilent 81689A, 81689B and 81649A Tunable Laser Modules Users Guide, Third Edition
2 Set the menu parameters to the values shown in Table 5.
3 Set the wavelength and power of your Tunable Laser module to the
values given in Table 6.
4 Press the key beside the laser output to switch on the laser output.
5 Wait until the wavelength meter has settled, then note the
wavelength displayed on the wavelength meter in the test record.
6 Increase the wavelength setting of the Tunable Laser module by the
steps shown in the test record.
7 Repeat steps 5 and 6 up to the maximum wavelength values shown
in Table 7.
8 Repeat steps 3 through 7 another 4 times.
9 From each repetition of the measurements, pick the maximum and
minimum deviations, and note these values in the test record.
10 Determine the Relative Wavelength Accuracy, Summary of all
Repetitions:
a Pick the largest Maximum Deviation, and note it as the Largest
Maximum Deviation in the test record.
b Pick t he sm allest Min imum Dev iation, and n ote it as t he Smallest
Minimum Deviation in the test record.
Table 5 Tunable Laser Channel Settings
Tunable Laser Channel Menu Parameters Values
<Wavelength Mode><λ>
<Source State><Off>
<Power Unit><dBm>
<Power Mode><Automatic>
Table 6 Initial Wavelength and Power Settings for Relative Wavelength Accuracy Tests
Module Wavelength] Power [P]
Agilent 81689A 1525.000 nm 3.00 dBm
Agilent 81689B 1525.000 nm 0.00 dBm
Agilent 81649A 1570.000 nm 3.00 dBm
Table 7 Maximum Wavelength for Relative Wavelength Accuracy Tests
Tunable Laser Module Maximum Wavelength Value
Agilent 81689A 1575 nm
Agilent 81689B 1575 nm
Agilent 81649A 1620 nm