Vista de página 76
4-44 TS-8900 Functional Test System Diagnostics User Guide
4 Diagnostic Testing Details
N9377A 16 Channel Load Card LC16CH_3 Tests
Display instruction to Install load module E2240-67012 on
LC16CH_3 Load Card.
Display instruction to install CEDGN Board #1 on TC1 and CEDGN
Board #4 on appropriate TC. Install Cable E6170-61607 from SLU
Utility Connector to J2 of Board #4. Install Cable E6170-61607 from
J1 of Board #1 to J1 of Board #4.
Setup DMM to perform 2W resistance measurement.
Measure LC16CH_3 load card loopback path resistance from
CH1-CH2 via Loadx.1 until CH7-CH8 via Loadx.1 using Row1 and
Row8
CH9-CH10 via Loadx.1 until CH15-CH16 via Loadx.1 using Row2 and
Row7
Measure LC16CH_3 load card sense resistor path:
CH1 Current Sense until CH16 Current Sense via Row25 and
Row26.
N9377A 16 Channel Load Card LC16CH_4 Tests
Display instruction to Install load module E2240-67012 on
LC16CH_4 Load Card.
Display instruction to install CEDGN Board #1 on TC1 and CEDGN
Board #4 on appropriate TC. Install Cable E6170-61607 from SLU
Utility Connector to J2 of Board #4. Install Cable E6170-61607 from
J1 of Board #1 to J1 of Board #4.
Setup DMM to perform 2W resistance measurement.
Measure LC16CH_4 load card loopback path resistance from
CH1-CH2 via Loadx.1 until CH7-CH8 via Loadx.1 using Row1 and
Row8
CH9-CH10 via Loadx.1 until CH15-CH16 via Loadx.1 using Row2 and
Row7
Measure LC16CH_4 load card sense resistor path:
CH1 Current Sense until CH16 Current Sense via Row25 and
Row26.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com