6
FUNCTION RANGE RESOLUTION
ACCURACY ± (% of reading + No. of Least Signifi cant Digit)
FREQUENCY
30 Hz ~ 45 Hz 45 Hz ~ 1 kHz 1 kHz ~ 10 kHz 10 kHz ~ 20 kHz 20 kHz~100 kHz
[1]
TRUE RMS AC
VOLTAGE
50.000 mV 0.001 mV 1.5+80 0.4+60 0.7+60 0.8+60 3.5+220
50.000 mV 0.01 mV
1.5+65
0.4+30 0.4+30
0.8+45 3.5+125
50.000 mV 0.1 mV
5.0000 V 0.0001 V
5.0000 V 0.001 V
500.00 V 0.01 V 1.5+45 3.5+125
[2]
500.00 V 0.1 V 0.4+45 0.4+45 1.5+45
[2]
N/A
U1252A AC+DC SPECIFICATIONS
FUNCTION RANGE RESOLUTION
ACCURACY ± (% of reading + No. of Least Signifi cant Digit)
FREQUENCY
30 Hz ~ 45 Hz 45 Hz ~ 1 kHz 1 kHz ~ 20 kHz
AC CURRENT
500.00 µA
[3]
0.01 µA
1.1+25 0.8+25 0.8+25
5000.0 µA 0.1 µA
50.000 mA 0.001 mA
1.2+25 0.9+25 0.9+25
440.00 mA 0.01 mA
5.0000 A 0.0001 A
1.8+30
[5]
0.9+30
3.3+70,
<3 A/5 kHz
10.000 A
[4]
0.001 A 0.9+25
RANGE RESOLUTION ACCURACY MINIMUM INPUT FREQUENCY
99.999 Hz 0.001 Hz
0.02%+3
<600 kHz
1 Hz
999.99 Hz 0.01 Hz
9.9999 kHz 0.0001 kHz
99.999 kHz 0.001 kHz
999.99 kHz 0.01 kHz
U1251A/U1252A FREQUENCY SPECIFICATIONS
[2]
[1] The additional error to be added as frequency >20 kHz and signal input <10% of range: 3 counts of LSD per kHz.
[2] The input signal is lower than the product of 20,000,000 V-Hz (product of voltage and frequency).
[3] Input current >35 mArms.
[4] Current can be measured from 2.5 A up to 10 A continuously. An additional 0.5% needs to be added to the specifi ed accuracy if the signal measured is in the range
of 10 A ~ 20 A for 30 seconds maximum. After measuring a current of >10 A, leave the meter to cool down for twice the measuring time used before application of
low current measurement.
[5] Input current < 3 Arms.
U1251A FREQUENCY SENSITIVITY DURING VOLTAGE MEASUREMENT
FREQUENCY SENSITIVITY AND TRIGGER LEVEL FOR U1251A
INPUT RANGE MINIMUM SENSITIVITY
(R.M.S. Sine-Wave)
TRIGGER LEVEL FOR DC COUPLING
(Maximum input for specifi ed
accuracy = 10 x Range or 1000 V)
20 Hz - 100 kHz >100 kHz ~ 200 kHz < 100 kHz >100 kHz ~ 200 kHz
50.000 mV 10 mV 15 mV 10 mV 15 mV
500.00 mV 25 mV 35 mV 60 mV 70 mV
1000.0 mV 40 mV 50 mV 100 mV 150 mV
5.0000 V 0.25 V 0.5 V 0.5 V / 1.25 V (< 100 Hz) 0.6 V
50.000 V 2.5 V 5 V 5 V 6 V
500.00 V 25 V N/A 50 V N/A
1000.0 V 50 V N/A 300 V N/A