SCPI Command Reference 6
N9340A User’s Guide 129
Types of Dtection
[:SENSe]:DETector:TRACe [1]|2|3|4
[FUNCtion]
AVERage|POSitive|SAMPle| NEGative|NORMal
[:SENSe]:DETector:TRACe [1]|2|3|4
[FUNCtion]?
This command specifies the detection mode. For
each trace interval (bucket), average detection
displays the average of all the samples within the
interval. The averaging can be done using two
methods:
•The power method (RMS)
The video method (Y Axis Units)
The method is controlled by the TRACE, Detector.
[:SENSe]:FREQuency Subsection
Center Frequency
[:SENSe]:FREQuency:CENTer <freq>
[:SENSe]:FREQuency:CENTer UP|DOWN
[:SENSe]:FREQuency:CENTer?
This command sets the center frequency.
*RST Value: 1.5 GHz
Available unit: Hz, kHz, MHz, GHz
Key Access: [FREQ] > {Center Freq}