Agilent Technologies N9320A Guia do Utilizador Página 143

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 182
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 142
One-Button Measurements 5
N9320A Spectrum Analyzer 135
a. Abs Stop is set to Couple and Fail Mask is set to Abs AND Rel.
a. Res BW is set to 100 kHz and Fail Mask is set to Rel.
Table 2 Offsets & Limits Defaulted for W-CDMA (3GPP) Mobile Station Measurements
a
Offset Start
Freq
(MHz)
Stop
Freq
(MHz)
Res BW
(kHz)
Abs
Start
(dBm)
Abs
Stop
(dBm)
Rel
Couple
Rel
Start
(dBc)
Rel Stop
(dBc)
Meas
BW
A. On 2.515 3.485 30.00 -71.07 -71.07 Man -35.23 -49.78 1
B. On 4.000 7.500 1000.0 -55.84 -55.84 Man -35.50 -39.00 1
C. On 7.500 8.500 1000.00 -55.84 -55.84 Man -39.00 -49.00 1
D. On 8.500 12.000 1000.00 -55.84 -55.84 Couple -49.00 -49.00 1
E. Off 12.500 15.000 1000.00 -55.84 -55.84 Couple -49.00 -49.00 1
Table 3 Offsets & Limits Defaulted for 802.11a and HiperLAN/2 Radio Standards
a
Offset Start
Freq
(MHz)
Stop
Freq
(MHz)
Abs
Start
(dBm)
Abs
Stop
(dBm)
Abs
Couple
Rel
Start
(dBc)
Rel Stop
(dBc)
Rel
Couple
Meas
BW
A. On 9.00 11.00 16.00 -4.00 Man 0.00 -20.00 Man 1
B. On 11.00 20.00 -4.00 -12.00 Man -20.00 -28.00 Man 1
C. On 20.00 30.00 -12.00 -24.00 Man -28.00 -40.00 Man 1
D. On 30.00 50.00 -24.00 -24.00 Couple -40.00 -40.00 Couple 1
E. Off 50.00 100.00 -24.00 -24.00 Couple -40.00 -40.00 Couple 1
Vista de página 142
1 2 ... 138 139 140 141 142 143 144 145 146 147 148 ... 181 182

Comentários a estes Manuais

Sem comentários