Save/Recall (Setups, Screens, Data) 16
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide 213
To save reference waveform files to a USB storage device
1 Press the [Save/Recall] key.
2 In the Save/Recall Menu, press the Save softkey.
3 In the Save Menu, press the Format softkey and turn the Entry knob to
select Reference Waveform data (*.h5) .
4 Press the Source softkey and turn the Entry knob to select the source
waveform.
5 Press the softkey in the second position and use the Entry knob to
navigate to the save location. See "To navigate storage locations" on
page 214.
6 Finally, press the Press to Save softkey.
A message indicating whether the save was successful is displayed.
To save masks
Mask files can be saved to one of four internal (\Agilent Flash) locations
or to an external USB storage device.
1 Press [Save/Recall] > Save > Format; then, turn the Entry knob to select
Mask (*.msk).
2 Press the softkey in the second position and use the Entry knob to
navigate to the save location. See "To navigate storage locations" on
page 214.
3 Finally, press the Press to Save softkey.
A message indicating whether the save was successful is displayed.
Mask files have the extension MSK.
See Also Chapter 14, “Mask Testing,” starting on page 187
NOTE
Masks are also saved as part of setup files. See "To save setup files" on page 207.