1 mW Power Reference
Note: The 1 mW power reference is provided for calibration of E-series and 8480 series sensors. The P-Series sensors are automatically
calibrated do not need this reference for calibration
Power output 1.00 mW (0.0 dBm). Factory set to ± 0.4% traceable to the National Physical Laboratory (NPL) UK
Accuracy (over 2-years) ±1.2% (0 to 55º C)
±0.4% (25 ± 10º C)
Frequency 50 MHz nominal
SWR 1.08 (0 to 55º C)
1.05 typical
Connector type Type N (f), 50
Rear panel inputs/outputs
Recorder output Analog 0-1 Volt, 1 k output impedance, BNC connector. For dual channel instruments there will be two
recorder outputs.
GPIB, 10/100BaseT LAN Interfaces allow communication with an external controller.
and USB2.0
Ground Binding post, accepts 4 mm plug or bare-wire connection
Trigger input Input has TTL compatible logic levels and uses a BNC connector.
Trigger output Output provides TTL compatible logic levels and uses a BNC connector
Line Power
Input voltage range 90 to 264 Vac, automatic selection
Input frequency range 47 to 63 Hz and 440 Hz
Power requirement N1911A not exceeding 50 VA (30 Watts)
N1912A not exceeding 75 VA (50 Watts)
Remote programming
Interface GPIB interface operates to IEEE 488.2 and IEC65.
10/100BaseT LAN interface.
USB 2.0 interface.
Command language SCPI standard interface commands.
GPIB compatibility SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP1, DC1, DT1, C0
Measurement speed
Measurement speed via remote interface 1500 readings per second
Regulatory information
Electromagnetic compatibility Complies with the requirements of the EMC Directive 89/336/EEC.
Product safety Conforms to the following product specifications:
EN61010-1: 2001/IEC 1010-1:2001/CSA C22.2 No. 1010-1:1993
IEC 60825-1:1993/A2:2001/IEC 60825-1:1993+A1:1997+A2:2001
Low Voltage Directive 72/23/EEC
Physical characteristics
Dimensions The following dimensions exclude front and rear panel protrusions: