13
Power meter Option 600
Display
Range: –100 dBm to +100 dBm
(range is power sensor dependent)
Display limits: ± 100 dBm (user settable)
Resolution: Settable 1.0, 0.1, 0.01, 0.001 in logarithmic mode,
or 1, 2, 3, or 4 significant digits in linear mode
Accuracy
Instrumentation:
Absolute: ± 0.02 dB (log) or ± 0.5 % (linear). Add the
corresponding power sensor linearity percentage.
Relative: ± 0.04 dB (log) or ± 1.0 % (linear). Add the
corresponding power sensor linearity percentage.
Zero set:
Zero set is the digital zero with a power sensor: ± 50 nW
Power reference:
Power output: 1.00 mW (0.0 dBm) traceable to the U.S. National
Institute of Standards and Technology (NIST).
Accuracy: ± 1.2 % worst case (± 0.9 % rss) for one year.
SWR: < 1.08
External attenuator
Max power: 100 W
Attenuation: 40 dB ± 0.5 dB
1
Power meter Option 600 with Agilent N8482A-CFT power sensor
(also supports Agilent 8481A/D and 8484A power sensors)
Frequency range: 100 KHz to 6 GHz
Dynamic Range: –30 dBm to +20 dBm
Measurement noise: < 114 nW
Zero drift: < ± 7 nW
EEPROM feature is disabled to maintain backward compatibility with 8480 Series
T1 analyzer Option 700
Features:
Receive level: (Line 1 and line 2) +6 dB DSC to –36 DB DSX or
100 mv p-to-p to 12 v p-to-p
Receive frequency display receive frequency
(5 ppm) (line 1 and line 2) “Loop-back” control
send CSU or NIU loop codes CSU/NIU emulation
respond to CSU or NIU loop codes
Electrical interface
Connectors, RX, TX: Primary and secondary ports
Output: Conforms to TR-TSY-000499, CCITT Rec.G.703
AT&T Pubs CB113, CB119, CB132, CB143
PUB62508 and PUB62411 pulse shape
specifications when terminated in 100 Ω and
0 dB line build-out is selected
Line build-out: 0 dB, –7.5 dB, –15 dB
Input:
Terminate: DSX +6 dB to DSX –36 dB, 100 Ω
Monitor: DSX –14 dB to DSX –40 dB, 100 Ω
Bridge: DSX +6 dB to DSX –36 dB, > 1000 Ω
Clock: 1.544 MHz
Internal: ± 5 ppm
External: ± 300 ppm
Recovered: ± 300 ppm
Transmitter and receiver
Framing: Unframed, D3/D4 & ESF
Channel formats: Full T1, 64x1
Test patterns: QRSS, all Os, 1:7, 2 in 8, 3 in 24, all 1s,
T-1-Daly, 55 OCTET
Error injection
Type: BPV, frame, CRC, pattern (logic)
Error rate: Single
Alarm inject
Type: LOS, LOF, yellow, AIS, idle (CDI)
E1 analyzer Option 710
Features
Error detect: Code (BPV), FAS, MFAS,CRC-4, far end block
(FEBE), pattern, frame slip
Error rate calculation: Bit-error-rate, error free seconds, errored seconds
Alarm detect: AIS, TS-16 AIS, FAST DISTANT, MFAS DISTANT
Clock and frame slips:
Clock slips, frame slips, peak wander, clock slip rate
Auto configuration: Automatically detect line code, framing and
test pattern
Receive level
(line 1 and line 2): +6 dB DSX to –36 dB DSX or 100 mv p-to-p
to 12 v p-to-p
Receive frequency (line1):
Display receive frequency (± 5 ppm)
Channel access: output audio to system
Delay measurement: Measure delay in unit intervals for
“looped-back” signal
Electrical interface
Connectors, RX, TX: Primary and secondary ports
Output: Conforms to ITU-T Rec.G.703
Line code: AMI, HDB3
Impedance:
Terminate: 75 Ω ± 5 % bridge: > 1000 Ω
Input: Terminate: DSX +6 dB to DSX –36 dB
Bridge: DSX +6 dB to DSX –36 dB
Clock: 2.048 MHz
Internal: ± 5 ppm
External: ± 300 ppm
Recovered: ± 300 ppm
Transmitter and receiver
Framing: Unframed, PCM-30, PCM-30 with CRC, PCM-31,
PCM-31 with CRC
Channel formats: Full E1, 64x1
Test patterns:
(True or Inverse, ITU Rec) 2
6
-1 (Q6 & Q5), 2
9
-1 (V.52),
2
11
-1 (0.152), 2
15
-1 (0.151) 2
20
-1 (V.57), QRSS,
2
23
-1 (0.151), all 0’s, 1:7, 1:3, 1:1, all 1’s
Error injection
Type: Code (BPV), FAS, MFAS, CRC-4, far end block
(FEBE), pattern
Error rate: Single
Alarm generation AIS, TS-16 AIS, FAS DISTANT,
MFAS DISTANT, loss of signal, loss of frame