Measurement Operations & Settings Measurement Functions – The SENSe Subsystem
110 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
command:
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:TERMination
syntax: :SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:TERMination
description: For the currently selected source, start the calibration and save the calibration values for a
defined termination reference measurement. See
“:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:REFLectance[l]” on page 112 for infor-
mation on setting the return loss value of your reference reflector.
parameters: none
response: none
example sens1:ret:cal:coll:term
affects: All return loss modules
command:
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:TERMination?
syntax: :SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:TERMination?
description: Queries the T-value (termination calibration value) for the return loss module
parameters: none
response: Termination calibration value as a float in dB
example sens1:ret:cal:term? +6.5000E+001
affects: All return loss modules
command:
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:VALues?
syntax: :SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:VALues?
description: Returns the the current calibration values
1. monitor diode reference power
2. return loss diode reference power
3. monitor diode parasitics power
4. return loss diode parasitics power.
parameters: Returns power values in W
response: none
example sens1:ret:cal:val
affects: All return loss modules