17
External DC bias input
1
Maximum voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .±40 V
Maximum current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20 mA
Measurement Basic Accuracy (Supplemental Performance Characteristics)
Measurement accuracy is specified at the connecting surface of the 7-mm
connector of the 43961A under the following conditions:
Warm-up time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .> 30 minutes
Ambient temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23 °C ±5 °C
(at the same temperature at which calibration was performed)
Signal level (@50 Ω terminated) . . . . . . . . . . . . . . . . . . . . . . . . . . . –6 to 14 dBm
Correction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ON
IFBW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .≤ 300 Hz
Averaging (cal) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ≥ 8
1. 2 kΩ ±5% resistor is inserted for DC bias current limitation.