Frequency response
2 ±5 °C, aft. = 10 dB, referenced to level at 20 MHz
10 MHz frequency 1.8 GHz . . . . . . . . . . . .< ±0.5 dB (±0.3 dB typically)
2 Hz frequency < 10 MHz . . . . . . . . . . . . . . .< ±1.5 dB (±0.8 dB typically)
For small signal measurement, fidelity is degraded by noise floor according to below formula:
20log
10
(1 ±10 x 3.5) dB typically
where x is signal to noise floor ratio in dB.
This fidelity error can be reduced by narrower video bandwidth or sweep averaging.
Linear scale
23 ±5 °C, –10 dBm [ ref level input att ] –50 dBm except for gain compression
RBW 300 kHz . . . . . . . . . . . . . . . . . . . . . . . . . . . .< ±3% of reference level
RBW 1 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . .< ±10% of reference level
IF gain switching uncertainty
input att. fixed, referenced to –20 dBm [ ref. level – input acct ] . . .< ±0.3 dB
Input attenuator switching uncertainty
20 dB to 40 dB, referenced to 10 dB . . . . . . . . . . . . . . . . . . . . . . . . .< ±1.0 dB
50 dB to 60 dB, referenced to 10 dB . . . . . . . . . . . . . . . . . . . . . . . . .< ±1.5 dB
RBW switching uncertainty
SPAN < 100 x RBW for RBW 10 kHz,
23 ±5 °C, referenced to 10 kHz RBW . . . . . . . . . . . . . . . . . . . . . . < ±0.5 dB
Temperature drift
S input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0.05 dB/°C typically
R, A, B inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.1 dB/°C typically
–x
20
Amplitude fidelity
Log scale
Amplitude fidelity Amplitude fidelity
1
Amplitude fidelity
1
Range @ 1 Hz RBW 3 kHz @ 10 kHz RBW 300 kHz @ 1 MHz RBW 3 MHz
(dB from ref. level) Spec. Typical Spec. Typical Spec. Typical
0 dB range –30 dB ±0.05 dB ±0.02 dB ±0.3 dB ±0.12 dB ±1.0 dB ±0.4 dB
–30 dB > range –40 dB ±0.01 dB ±0.03 dB ±0.3 dB ±0.12 dB ±1.0 dB ±0.4 dB
–40 dB > range –50 dB ±0.12 dB ±0.05 dB ±0.4 dB ±0.15 dB ±1.2 dB ±0.5 dB
–50 dB > range –60 dB ±0.4 dB ±0.12 dB ±0.1 dB ±0.3 dB ±1.4 dB ±0.6 dB
–60 dB > range –l0 dB ±1.2 dB ±0.8 dB ±1.5 dB ±0.6 dB ±2.2 dB ±0.8 dB
–10 dB > range –80 dB ±4 dB ±1 dB ±4.3 dB ±1.2 dB
–80 dB > range –90 dB +3 dB
–90 dB > range –100 dB ±10 dB
1. At 23 ±5 °C, 10 dBm [ ref. level input att ] > –50 dBm except for gain compression