U1401B Quick Start Guide
Performing Resistance Measurement and
Continuity Test
CAUTIONCAUTION
Disconnect circuit power and discharge all high-voltage
capacitors before measuring resistance to prevent possible
damage to the instrument or the device under test.
1 Set the rotary switch to .
2 Connect the red (+) and black (–) test
leads to the INPUT terminals.
3 Probe the resistor (or shunt) leads and
read the display.
4 To perform continuity test, press
to toggle the audible continuity function
on or off.
For continuity test, the instrument will beep if
the resistance is less than the values
indicated below:
Measurement range Resistance threshold
500.00 Ω 10 Ω
5.0000 kΩ 100 Ω
50.000 kΩ 1 kΩ
500.00 kΩ 10 kΩ
5.0000 MΩ 100 kΩ
50.000 MΩ 1 MΩ