140 Agilent U1251B/U1252B User’s and Service Guide
6 Performance Tests and Calibration
Using the front panel for adjustments
This section describes the procedures to perform
adjustments from the front panel.
Selecting the Adjustment Mode
To unsecure the instrument, see “Unsecuring the instrument
for calibration on page 136 or “To unsecure the instrument
without the security code” on page 138. Once unsecured, the
reference value will be indicated on the primary display.
Entering Adjustment Values
Use the following adjustment procedure to enter an input
calibration value from the front panel:
1 Use the edit keys and to select each digit in
the primary display.
2 Use the and the arrow keys to advance
through the digits 0 through 9.
3 Press when done to start calibration.