4
Measurement Accuracy: ±20 mrad using a measurement interval
of 12.5 ms
Measurement Resolution: 0.01 mrad
Signal Generator Specifications
RF Frequency
Range: 400 kHz to 1000 MHz, and 1.7 to 2.0 GHz
Accuracy and Stability: ±(0.065 Hz plus reference oscillator
accuracy)
Output
RF IN/OUT Connector
Level Range: –137 to –40 dBm into 50 W
Level Accuracy: ±1.0 dB (level >–127 dBm); if RF Analyzer
is also connected add ±0.1 dB
Typically ±1.0 dB for levels below –127 dBm
Reverse Power: 100 watts continuous, temperature <40° C,
75 W continuous, for temperature <55° C
SWR: <1.5:1
DUPLEX OUT Connector
Level Range: –125 to –10 dBm into 50 Ω
Maximum Power Output: +3 dBm
Level Accuracy: ±1.5 dB, Typically ±1.0 dB for all levels
Reverse Power: 200 mW max
SWR: <1.7:1
Supplemental Characteristics
Resolution: 0.1 dB
Spectral Purity
(For output levels of <–10 dBm at DUPLEX OUT or <–40 dBm
at RF IN/OUT)
Harmonics: <–25 dBc
Non-Harmonic Spurious (>5 kHz from carrier):
250 kHz ≤f
c
<249 MHz <–45 dBc
249 MHz ≤f
c
≤1000 MHz <–60 dBc
1700 MHz ≤f
c
≤2000 MHz <–55 dBc
Residual FM (rms, CCITT filter):
250 kHz ≤f
c
<249 MHz <7 Hz
249 MHz ≤f
c
<501 MHz <4 Hz
501 MHz ≤f
c
≤1000 MHz <7 Hz
1700 MHz ≤f
c
≤2000 MHz <14 Hz
Supplemental Characteristics
SSB Phase Noise (20 kHz offset):
f
c
<1 GHz <–116 dBc/Hz
1.7 GHz < f
c
<2.0 GHz <–90 dBc/Hz
FM
FM Deviation Maximum (For rates >25 Hz):
400 kHz <f
c
<249 MHz 100 kHz
249 MHz <f
c
<501 MHz 50 kHz
501 MHz <f
c
<1000 MHz 100 kHz
1.7 GHz <f
c
<2.0 GHz 100 kHz
[FM not specified for (f
c
minus FM dev) <400 kHz]
FM Rate (1 kHz reference):
Internal: DC to 25 kHz (1 dB BW )
External: AC Coupled: 20 to 75 kHz (typically 3 dB BW)
DC Coupled: DC to 75 kHz (typically 3 dB BW)
FM Accuracy (1 kHz rate):
<10 kHz dev: ±3.5% of setting ±50 Hz
>10 kHz dev: ±3.5% of setting ±500 Hz
FM Distortion (THD + noise, in a 0.3 to 3 kHz BW):
<0.5% at >3 kHz dev. and 1 kHz rate, f
c
<1000 MHz
<1.0% at >3 kHz dev. and 1 kHz rate, 1.7 GHz < f
c
<2.0 GHz
Center Frequency Accuracy in DC FM mode
(external source impedance <1k Ω):
±500 Hz (after DC FM zero), typically ±50 Hz
Supplemental Characteristics
Ext. Mod. Input Impedance: 600 Ω nominal
Resolution:
50 Hz for <10 kHz deviation
500 Hz for >10 kHz deviation
AM
Supplemental Characteristics
AM Depth: 0 to 60%
External Modulation Input Impedance:
600 Ω nominal
RF Analyzer Measurements
SWR:
RF IN/OUT Port: < 1.5: 1
ANT IN Port: <1.6:1, except <1.8:1 for 0 dB
attenuation in 1.7 to 2.0 GHz band
RF Frequency Measurements
Measurement Range: 400 kHz to 1 GHz, 1.7 to 2.0 GHz
Level Range:
RF IN/OUT:
1 mW to 100 watts continuous, temp. <40° C,
75 W continuous, for temperatures <55° C
ANT IN: –38 dBm to +15 dBm
Accuracy: ±1 Hz plus timebase accuracy
Supplemental Characteristics
Frequency Resolution: 1 Hz (The user must set the instrument
within 15 kHz of the signal under test)
Comentários a estes Manuais