5
AM Measurement
Frequency Range: 10 MHz to 1 GHz (usable to 400 kHz), 1.7 GHz
to 2.0 GHz
Depth: 0 to 95%
Input Level Range (levels in PEP):
–6 dBm to +52 dBm at RF IN/OUT
–44 dBm to +14 dBm at ANT IN
Supplemental Characteristics
Accuracy: ±5% of reading ±1.5% AM (50 Hz to 10 kHz rates,
modulation <80%)
THD + Noise: <2% rms for modulation <80% AM
(at 1 kHz rate in a 0.3 to 3 kHz bandwidth)
Residual AM: <0.2% in a 0.3 to 3 kHz bandwidth
Resolution: 0.1%
Spectrum Analyzer
Frequency Range: 400 kHz to 1 GHz, 1.4 GHz to
2.0 GHz
Frequency Span/Resolution Bandwidth (coupled)
Span Bandwidth
<50 kHz 300 Hz
<200 kHz 1 kHz
<1.5 MHz 3 kHz
<18 MHz 30 kHz
>18 MHz 300 kHz
Span Capability:
2
Full span capability 1 GHz
Display: Log with 1, 2, and 10 dB per division
Display Range: 80 dB or 8 divisions
Reference Level Range: +50 dBm to –50 dBm
Residual Responses: <–70 dBm (ANT IN port, no input
signal, 0 dB attenuation)
Image rejection: >50 dB
Supplemental Characteristics
Non-Harmonic Spurious Responses:
>65 dB down (for input signals ≥50 MHz, <–30 dBm),
>55 dB down (for input signals <50 MHz, <–30 dBm).
Log Scale Linearity: ±2 dB (for input levels
<–30 dBm within 6 divisions of reference level
Displayed Average Noise Level: <–114 dBm, for
<50 kHz span, –50 dBm reference level
Level Accuracy (at center frequency and within
one division of reference level): ±2.5 dB
2. Center frequency must be <1.0 GHz or between 1.7 GHz to 2.0 GHz. Portions of
the 1 GHz span which are outside of the analyzer’s frequency range are truncated.
Tracking Generator
Tracking Generator Frequency Range: 400 kHz to 1 GHz,
1.7 GHz to 2.0 GHz
Frequency Offset: Frequency span endpoints ± frequency offset
cannot be <400 kHz or >1 GHz in the 1 GHz band and cannot be
<1.7 GHz or >2.0 GHz in the 2 GHz band.
Output Level Range: Same as signal generator on page 3.
Sweep Modes: Normal and inverted
Oscilloscope Specifications
Frequency Range: 2 Hz to 50 kHz (3 dB BW)
Scale/Division: 10 mV to 10 V
Amplitude Accuracy: ±1.5 % of reading ±0.1 division (20 Hz to
10 kHz)
Time/Division: 10 µsec to 100 msec
Trigger Delay: 20 µsec to 3.2 seconds
Supplemental Characteristics
3 dB Bandwidth: Typically >100 kHz
Internal DC Offset: <0.1 div (>50 µV/div sensitivity)
AF Analyzer Specifications
Frequency Measurement
Measurement Range: 20 Hz to 400 kHz
Accuracy: ±0.02% plus 0.1 Hz plus timebase accuracy
External Input: 20 mV to 30 Vrms
Supplemental Characteristics
Resolution: 0.01 Hz, f <10 kHz; 0.1 Hz, f <100 kHz; and
1 Hz for f >100 kHz
AC Voltage Measurement
Measurement Range: 0 to 30 Vrms
Accuracy: ±3% of reading (20 Hz to 15 kHz, inputs >1 mV)
Residual Noise: 150 µV (15 kHz LPF) 450 µV (>99 kHz LPF)
Supplemental Characteristics
3 dB Bandwidth: Typically 2 Hz to 100 kHz
Nominal Input Impedance: Switchable between 1 MΩ in parallel
with 95 pF or 600 Ω floating
Minimum Resolution: 4 digits for inputs >100 mV; 3 digits for
inputs <100 mV