7
Z-Axis blanking/markers output
Supplies positive rectangular pulse (approximately
+5 volts into 2 kohms) during the retrace and band
switchpoints of the RF output. Also supplies a neg-
ative pulse (–5 volts) when the RF is at a marker
frequency (intensity markers only). (BNC female,
rear panel.)
Volts/GHz output
Supplies a voltage proportional to output frequency
at 0.25 volts/GHz, 0.5 volts/GHz, or 1 volt/GHz
(model dependant and internally switchable).
Maximum output 18 volts. Minimum load imped-
ance 2 kohms. Accuracy ±0.5%, ±10 mv, typical.
(BNC female, rear panel.)
Source module interface
Provides bias, flatness correction, and leveling con-
nections for the 83550 series of millimeter-wave
source modules. (Special, front, and rear panels.)
Auxiliary interface
Provides control signal connections to the 8516A
S-parameter test set. Also used when two 8360
series synthesized sweepers are operated in mas-
ter/slave mode. (25-pin D-subminiature receptacle,
rear panel.)
Pulse video output (Option 002 only.) Outputs the
pulse modulation waveform that is supplied to the
modulator. This can be either the internally or
externally generated pulse modulation signal.
(BNC female, rear panel.)
Pulse sync out (Option 002 only.) Outputs a 50 ns
wide TTL pulse synchronized to the leading edge of
the internally generated pulse. (BNC female, rear
panel.)
AM/FM output (Option 002 only.) Outputs the inter-
nally generated AM or FM waveform. This output
can drive 50 ohms or greater. The AM output is
scaled the same as it is generated, either 100%/V or
10 dB/V. The FM scaling depends on the FM devia-
tion selected. (BNC female, rear panel.)
Models
83620B 10 MHz to 20 GHz
83622B 2 GHz to 20 GHz
83623B 10 MHz to 20 GHz (high power)
83624B 2 GHz to 20 GHz (high power)
83630B 10 MHz to 26.5 GHz
83640B 10 MHz to 40 GHz
83650B 10 MHz to 50 GHz
83623L 10 MHz to 20 GHz
83630L 10 MHz to 26.5 GHz
83640L 10 MHz to 40 GHz
83650L 10 MHz to 50 GHz
Options
Option 001 adds step attenuator
With this option, minimum settable output power
is –110 dBm. Maximum leveled output power is
reduced by 1.5 dB to 20 GHz, 2 dB above 20 GHz,
and 2.5 dB above 40 GHz.
Option 002 adds internal modulation generator
(8360B only – not available on 8360L)
Adds a digitally synthesized modulation waveform
source-on-a-card to the 8360. Provides signals that
would otherwise be applied to the external modu-
lation inputs.
Option 004 rear panel RF output
Moves RF output, external ALC input, pulse
input/output, AM input, and FM input connectors
to the rear panel.
Option 006 fast pulse modulation
(8360B only – not available on 8360L)
Improves pulse rise/fall time to 10 ns. Also
improves harmonic performance.
Option 008 1 Hz frequency resolution
Provides frequency resolution of 1 Hz.
Option 700 MATE system compatibility
Provides CIIL programming commands for MATE
system compatibility.
Option 806 rack slide kit
Used to rack mount the 8360 while permitting
access to internal spaces.
Option 908 rack flange kit
Used to rack mount the 8360 without front handles.
Option 910 extra operating and service manuals
Provides a second copy of Operating and Service
manuals.
Option 913 rack flange kit
Used to rack mount the 8360 with front handles.
Front handles are standard on the 8360.
ISO 9002 compliant
These models are manufactured in an ISO 9002
registered facility in concurrence with Agilent
Technologies’ commitment to quality.
Upgrades
Model and frequency upgrades are available.
Please contact your Agilent sales representative
for details.