8510C Service Quick Reference Guide 3-3
8510C Adjustments
Procedure 1. Vertical Alignment Adjustment (CRT)
Procedure 1. Vertical Alignment Adjustment (CRT)
The original 85101C display/processor incorporated a cathode ray tube (CRT). The current
design incorporates a liquid crystal display (LCD). This procedure applies only to
instruments that are equipped with a CRT.
NOTE The vertical alignment can be adversely affected by magnetic interference.
Before adjusting the vertical position, be sure the analyzer is in a
non-magnetic environment and the CRT is degaussed.
1. Switch on the system and allow it to warm up for 60 minutes.
2. To access the vertical position and focus controls, remove the side panel nearest the
display.
3. Insert a narrow, non-conductive, flat-head screwdriver ( 2 or more inches long) into the
vertical position hole.
4. Adjust the control until the softkey labels align with the softkeys.
Figure 3-1 Vertical Adjustment Location