Table 60503-4. Calibration Information
Ranges and
Calibration Points
Variables Variables
Value
Power Supply
Settings
Current
Shunt
High Current Range Hi_curr_rng 10 25 V/10.5 A 15 A
High Current Offset Hi_curr_offset 0.0048
Low Current Range Lo_curr_rng 1 25 V/2 A 15 A
Low Current Offset Lo_curr_offset 0.0032
Voltage Range N/A N/A 246 V/0.6 A N/A
Voltage Hi point Volt_hipt 240
Voltage Lo point Volt_lopt 2
Low Resistance Range Lo_res_rng 24 60 V/1.82 A 15 A
Low Resistance Hi point Lo_res_hipt 23.9
Low Resistance Lo point Lo_res_lopt 0.88
Middle Resistance Range Mid_res_rng 240 43.6 V/4 A 15 A
Middle Resistance Hi point Mid_res_hipt 500
Middle Resistance Lo point Mid_res_lopt 24
High Resistance Range Hi_res_rng 24020 240 V/2 A 15 A
High Resistance Hi point Hi_res_hipt 2000
High Resistance Lo point Hi_res_lopt 240