Time interval resolution
and accuracy
One of the key advances the 53200
Series provides is the industry
leading timing specifications. The
single-shot resolution for the 53230A
is 20 picoseconds while on the previ-
ous 531xxA generation, the fastest
single-shot timing available was
150 ps (53132A). Single-shot timing
represents how well the counter can
resolve an event in time. It is most
often associated with time interval
measurements, but every counter
measurement is a result of timing and
accurate placement of the edge. So,
the better the single-shot timing of a
frequency counter, the better overall
measurement performance of the
counter.
Comparison Summary
Agilent 531xxA Series
(53181A, 53131A, 53132A)
Agilent 53200 Series
(53210A, 53220A, 53230A)
Frequency bandwidth 225 MHz, 1.5 (53181A only), 3, 5, 12.4 GHz 350 MHz, 6, 15 GHz
Frequency resolution 10 digits/s (53181A, 53131A)
12 digits/s (53132A)
10 digits/s (53210A)
12 digits/s (53220A, 53230A)
Time interval resolution
(single-shot resolution)
500 ps (53131A)
150 ps (53132A)
100 ps (53220A)
20 ps (53230A)
Memory None 1M reading internal memory
USB flash
Speed Transactional: 200 readings/s
Readings to memory: N/A
Timestamp: N/A
Transactional: 350 - 400 readings/s
Readings to memory:
Frequency: up to 75,000 readings/s
Time interval: up to 90,000 readings/s
Timestamp: up to 1M readings/s
Timebase (See detailed
comparison below)
XO (standard in all models)
3 OCXO options
TCXO (standard in all models)
1 OCXO Option
Connectivity GPIB, RS232 (print only) LXI-C/LAN, USB, GPIB
Display & analysis Vacuum florescent display
Math and limit operations
Color graphical display
Full integrated math, limits, statistics,
graphs
Rack mount dimensions 212.6 mm (W) x 88.5 mm (H) x 348.3 mm (D)
(2U x ½ width)
212.8 mm (W) x 88.3 mm (H) x
272.3 mm (D) (2U x ½ width)
Memory and measurement
reading speed
There are several ways to consider
the measurement speed on frequency
counters:
Single measurement reading
throughput: the time to take a single
measurement and transfer from vola-
tile reading memory over the I/O bus.
Block reading throughput: the time
to take blocks of measurements into
memory and when complete, transfer
from volatile reading memory over the
I/O bus.
Typically, single measurement
throughput is slower than block
measurement throughput due to I/O
processing/transfer overhead. The
531xxA frequency counters are single
reading counters with no internal
memory. In most applications, up to
200 readings per second is the typical
single measurement reading speed
on 531xxA counters. Depending on
the I/O interface chosen (USB, LAN
or GPIB), the single measurement
throughput on the 53200 Series coun-
ters can have optimized throughput
of approximately 400 readings per
second.
The reading throughput potential is
substantially different when block
reading capabilities are considered.
With the built-in memory of 1 mil-
lion readings on the 53200 Series
counters, speeds increase to tens of
thousands of readings per second.
Again, the I/O interface chosen
impacts the amount of overhead on
the overall throughput capabilities.
The 53200 Series data sheet outlines
typical reading speeds by model (part
number 5990-6283EN).