Service Guide N5230-90025 3-9
PNA Series Microwave Network Analyzers Tests and Adjustments
N5230C Non-ANSI/NCSL Z540–1–1994 Verification
Non-ANSI/NCSL Z540–1–1994 Verification
For non-ANSI/NCSL Z540-1-1994, perform the preliminary checks and performance tests while stopping to
troubleshoot. Refer to Figure 3-2 for test flow. Troubleshoot and repair the first problem encountered without
continuing to other tests. After you troubleshoot, repair, and adjust, repeat the last failed portion and
generate a new set of data.
Figure 3-2 Non–ANSI/NCSL Z540–1–1994 Test Path Verification Flowchart