Agilent Technologies InfiniiVision 7000B Series Guia do Utilizador Página 242

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 241
6 Measurements and Math Functions
242 InfiniiVision Oscilloscopes User’s Guide
Enabling Precision Measurements and Math
Normally, after pressing [Save/Recall] > Default Setup, the oscilloscope performs
measurements and generates math waveforms using a 1000-point (maximum)
measurement record. This measurement record is purposely small in order to provide
high waveform update rates and minimal “dead time” between acquisitions to improve
the probability of capturing infrequent events.
However, at the expense of waveform update rate, you can enable precision
measurements and math waveforms that use a 128K-point (maximum) record. This mode
gives you better resolution on measurements and math waveforms (including FFT).
To enable precision measurements and math:
1 Press [Utility] > Options.
2 In the User Options Menu, select Precision to enable precision measurements and
math waveforms.
Precision measurements and math are disabled in the same way, or you can disable them
by pressing [Save/Recall] > Default Setup.
Vista de página 241
1 2 ... 237 238 239 240 241 242 243 244 245 246 247 ... 427 428

Comentários a estes Manuais

Sem comentários