4 Triggering
104 InfiniiVision Oscilloscopes User’s Guide
When the duration trigger occurs
The timer starts on the last edge that makes the pattern (logical AND) true. The trigger will occur
on the first edge that makes the pattern false if the time qualifier criteria of the pattern has been
met, except in Timeout mode. In Timeout, the trigger occurs when the timeout value is reached
while the pattern is true.