11 Mask Test
338 InfiniiVision Oscilloscopes User’s Guide
Mask Test Trigger Output
The oscilloscope’s trigger output (located on the rear panel) can be activated by mask
test results.
Press [Utility]&Options&Rear Panel&Trig Out and select Mask. Use the Mask Out
softkey to select the desired mode.
The Pass/fail status is evaluated periodically. When the evaluation of the testing period
results in a pulse output, the trigger output pulses high (+ 5 V). Otherwise, the trigger
output remains at low (0 V).
Run mode Mask Out setting Trigger Output (rear panel)
Single Pass A pulse is produced if the mask test
passed.
Single Fail A pulse is produced if the mask test
failed.
Run Pass A pulse is produced if all mask tests
have passed
Run Fail A pulse is produced if any mask test
has failed