Agilent EasyEXPERT Users Guide Vol. 2, Edition 1 6- 15
Remote Control Interface
BENCh Subsystem
[:BENCh][:SELected]:RUN[:SINGle]
Starts the single measurement of the test now opened. Use the *OPC? command to
confirm the completion of the single measurement.
Syntax [:BENCh][:SELected]:RUN[:SINGle]
Example :RUN
[:BENCh][:SELected]:STRing
Sets or returns the value of the specified string parameter defined in the application
test now opened.
Syntax [:BENCh][:SELected]:STRing "param_name", "svalue"
[:BENCh][:SELected]:STRing? "param_name"
Parameter param_name Parameter name. Data type is SPD.
svalue Value of the parameter. Data type is SPD.
Query response "svalue"<newline>
Data type is SRD.
Example :STR "Base", "SMU1:HP"
:STR? "Base"