Agilent Technologies EasyEXPERT Guia do Utilizador Página 105

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 228
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 104
Agilent EasyEXPERT Users Guide Vol. 2, Edition 1 8-29
Application Library
QSCV Maximum Measurement Value and Accuracy
Figure 8-9 Calculation Example of Measurement Accuracy: HRSMU, ASU+SMU
Conditions:
QSCV Operating Mode: Normal
Measurement Range: 100 pA
Output Range: 20 V
Integration Time: 2, 1, 0.5, 0.3 s corresponding to the lines from top to bottom
Equivalent Parallel Resistance of DUT: 100 G
Guard Capacitance of Measurement Path: 200 pF
0
2
4
6
8
10
12
14
0.001 0.01 0.1 1 10
QSCV Measurement Voltage (V)
Reading Accuracy (%)
1.E-14
1.E-13
1.E-12
1.E-11
1.E-10
0.001 0.01 0.1 1 10
QSCV Measurement Voltage (V)
Offset Accuracy (F)
Vista de página 104
1 2 ... 100 101 102 103 104 105 106 107 108 109 110 ... 227 228

Comentários a estes Manuais

Sem comentários