Ch 93 GU
Measurement 1. Make Hardware Connections
Turn mainframe power OFF
Connect DMM leads as shown in Figure 2-8
Turn mainframe power ON
2. Measure Channel 93 GU Resistance
Send *RST to multiplexer
Send CLOS (@nn93) to close chan 93, where nn = card #
Trigger the DMM with TRIG SGL and note reading
Send OPEN (@nn93) to open channel 93
Subtract measured value of (R40 + R41) from DMM reading
Enter the result in Table 2-2 for Channel 93 GU
Figure 2-8. Ch 93 GU Measurement Connections