Agilent Technologies E1343A Manual de Serviço

Consulte online ou descarregue Manual de Serviço para Medição, teste e controlo Agilent Technologies E1343A. Agilent Technologies E1343A Service manual [en] Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 0
Agilent 75000 Series B
Agilent E1343A, E1344A, E1345A, E1347A,
E1355A and E1356A
Relay Multiplexers
Service Manual
Copyright © Agilent Technologies, Inc. 2005
Printed in Malaysia
Vista de página 0
1 2 3 4 5 6 ... 83 84

Resumo do Conteúdo

Página 1 - Relay Multiplexers

Agilent 75000 Series BAgilent E1343A, E1344A, E1345A, E1347A,E1355A and E1356ARelay MultiplexersService Manual

Página 3 - Contents

What’s in This ManualManual Overview This manual shows how to service the E1343A, E1344A, E1345A, E1347A,E1355A, and E1356A Relay Multiplexers. Consul

Página 4

Notes10 What’s In This Manual E1343A/44A/45A/47A/55A/56A Service Manual

Página 5

Chapter 1General InformationIntroductionThis manual contains information required to test, troubleshoot, and repairthe following Agilent relay multip

Página 6 - Safety Symbols

Safety ConsiderationsThese products are Safety Class I instruments provided with a protectiveearth terminal when installed in the mainframe. Check th

Página 7 - Declaration of Conformity

WARNING USING AUTOTRANSFORMERS. If the mainframe is to beenergized via an autotransformer (for voltage reduction) makesure the common terminal is conn

Página 8

Relay LifeElectromagnetic relays are subject to normal wear-out. Relay life dependson several factors. Two factors are loading and switching frequency

Página 9

ReplacementStrategyThe replacement strategy also depends on the application. If some relays areused more often, or at higher load, than the others, th

Página 10

MultiplexerSpecificationsRefer to the individual module’s User’s Manual for specifications. Thesespecifications are the performance standards or limit

Página 11 - What’s in This Manual

Recommended Test EquipmentTable 1-1 lists the test equipment recommended for testing, adjusting, andservicing the relay multiplexers. Essential requir

Página 13 - General Information

Figure 1-2. Initial (Incoming) Inspection Guidelines18 General Information Chapter 1

Página 14 - Safety Considerations

Shipping Guidelines Follow the steps in Figure 1-3 to return a relay multiplexer to an AgilentTechnologies Sales and Service Office.1. Prepare the Mu

Página 15

Notes20 General Information Chapter 1

Página 16 - Relay Life

Chapter 2Verification TestsIntroductionThis chapter describes the verification tests for the E1343A, E1344A,E1345A, E1347A, E1355A, E1356A Relay Mult

Página 17 - Multiplexer Descriptions

Functional Verification TestThe Functional Verification Test for the multiplexer modules consists ofsending the *IDN? command and checking the respons

Página 18

Performance Verification TestsThe procedures in this section are used to test the multiplexer’s electricalperformance using the specifications as the

Página 19 - Inspection/Shipping

Test 2-1: Closed Channel Resistance TestThis test verifies that all relay contacts meet the closed-channel resistancespecification for the multiplexer

Página 20

Chs 00 - 07 and 90HI Measurements 1. Make Hardware Connections– Turn mainframe power OFF– Connect DMM leads as shown in Figure 2-3– Turn mainframe po

Página 21

3. Repeat for Channels 01 - 07 LO and Channel 90 LO– Repeat steps 1 and 2 for channels 01 - 07 LO and 90 LO– Use CLOS (@nncc) and OPEN (@nncc), where

Página 22

3. Repeat for Channels 01 - 07 GU and Channel 90 GU– Repeat steps 1 and 2 for channels 01 - 07 GU and 90 GU– Use CLOS (@nncc) and OPEN (@nncc), where

Página 23 - Verification Tests

Contents E1343A/44A/45A/47A/55A/56A Service ManualWarranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 24 - Operation Verification Test

2. Measure Channel 08 LO Resistance– Send *RST to multiplexer– Send CLOS (@nn08) to close chan 08, where nn = card #– Trigger the DMM with TRIG SGL a

Página 25 - Wiring the Test

2. Measure Channel 08 GU Resistance– Send *RST to multiplexer– Send CLOS (@nn08) to close chan 08, where nn = card #– Trigger the DMM with TRIG SGL a

Página 26 - Resistors

200 PRINT TABXY(1,5),"Measure resistor R";I;"value (in Ohms)"210 INPUT " Enter resistor value (in Ohms), then press C

Página 27 - LO Measurements

600 PRINT "Resistance for Channel";I;"Relay is >2.0 Ohms"610 PAUSE620 END IF630 NEXT I640 IF K<

Página 28 - GU Measurements

990 IF I=8 AND K=2 THEN Result1(I,K)=Value1(I,K)-(R(25)+R(31))1000 IF I=9 AND K=0 THEN Result1(I,K)=Value1(I,K)-(R(23)+R(26))1010 IF I

Página 29 - HI Measurements

Typical ResultsChannel HI (OHMS) LO (OHMS) GU (OHMS)0 1 2 3 4 5 6 790 8 910111213141591920.510.490.480.420.520.520.490.460.490.470.510.430.450.500.480

Página 30

HI to Chassis Isolation 1. Make hardware connections as shown in Figure 2-92. Set DMM to 2-wire ohms, 1 GΩ range3. Send CLOS (@nn00:nn15, nn90:nn93

Página 31 - Channel Resistance

HI to GU Isolation1. Make hardware connections as shown in Figure 2-112. Trigger the DMM with TRIG SGL 3. Record the DMM reading on Table 2-2 (HI t

Página 32 - Continued on Next Page

160 PAUSE 170 CLEAR SCREEN 180 ! 190 ! Measure DC Isolation (HI to Chassis, HI to LO, HI to GU) 200 ! 210 OUTPUT @Mux;"*RST" 22

Página 33

Performance Test RecordTable 2-2, "Performance Test Record," is a form you may copy and use torecord performance verification test results f

Página 34

Chapter 3. Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41Introduction

Página 35 - Test 2-2: DC Isolation Test

Table 2-2. Performance Test Record (Page 1 of 3)General InformationTest Facility:Name_______________________________________Address____________

Página 36 - HI to LO Isolation

Table 2-2. Performance Test Record (Page 2 of 3)Test No/Description Minimum*ValueMeasured Value (V)MaximumValueMeasurementUncert Test AccRatio (TAR)2

Página 37 - Example: DC Isolation

Table 2-2. Performance Test Record (Page 3 of 3)TestNo/DescriptionMinimum*ValueMeasured Value (V) Maximum** ValueMeasUncert Test AccRatio(TAR)2-1. Clo

Página 38

Chapter 3Replaceable PartsIntroductionThis chapter contains information on ordering replaceable parts for theE1343A/44A/45A/47A/55A/56A Relay Multipl

Página 39 - (0.005 ⋅ 1.2×10

Table 3-1. Relay Multiplexer Assembly/Terminal Module Part NumbersMultiplexer ComponentAssemblyTerminalModuleTerminalCardTerminalCaseAssyE1343A 16-Cha

Página 40

Table 3-2. E1345A/47A/55A/56A Replaceable Parts (cont’d)ReferenceDesignatorPartNumberQty Part DescriptionTERMINAL MODULESE1345-80001 1 E1345A TERMINA

Página 41

Table 3-3. E1343A/44A Replaceable Parts (see applicable serial number ranges on first page of this chapter)ReferenceDesignatorPartNumberQty Part D

Página 42

Table 3-4. Reference Designators Reference DesignatorsA assembly PCB printed circuit boardBRK bracket PNL panel C capacitor Q transistorCR diode R res

Página 43 - Replaceable Parts

Parts LocatorsFigures 3-1 through 3-3 show locations of replaceable parts for the E1343A,E1344A, E1345A, E1347A, E1355A, and E1356A Relay Multiplexers

Página 44

Figure 3-2. E1343A, E1344A, E1345A, E1347A Terminal Module Mechanical PartsChapter 3 Replaceable Parts 47

Página 45 - (cont’d)

CertificationAgilent Technologies, Inc. certifies that this product met its published specifications at the time of shipment from the factory. Agilent

Página 46

Figure 3-3. E1355A/E1356A Terminal Module Mechanical Parts48 Replaceable Parts Chapter 3

Página 47 - Reference Designators

Chapter 4ServiceIntroductionThis chapter contains generic service information for the multiplexers.Also included are troubleshooting, repair, and mai

Página 48 - Parts Locators

E1343A/44AMultiplexerDescriptionsThe E1343A 16-Channel High Voltage Relay Multiplexer and the E1344A16-Channel High Voltage Relay Multiplexer (with

Página 49

Terminal module connections include H, L, and G terminals for Channels00 - 15, Bank 0 and Bank 1 Commons, and the AT and BT Tree Switches.In addition,

Página 50

Figure 4-1. E1343A, E1344A, E1345A, E1347A Multiplexer Block Diagram52 Service Chapter 4

Página 51 - Introduction

Terminal ModuleDescriptionThe terminal module provides connections from user devices to themultiplexer. The terminal module also provides connections

Página 52 - Descriptions

Figure 4-2. E1355A/56A Multiplexers Block Diagram54 Service Chapter 4

Página 53 - E1345A/47A

Table 4-1. E1355A/E1356A Relay FunctionsRelay Function NotesK100K101K102K103Measure strain (channel 00)Measure strain (channel 01) Measure strain (cha

Página 54 - 52 Service Chapter 4

Troubleshooting TechniquesTo troubleshoot a relay multiplexer problem, you must first identify theproblem and then isolate the cause of the problem to

Página 55 - Gage Multiplexer

Making VisualChecksVisual checks for the multiplexer assemblies include the following. See Table 4-3 for typical symptoms/actions. • Check switches/ju

Página 56 - 54 Service Chapter 4

Frame or chassis ground terminal—typi-cally connects to the equipment’s metalframe.Alternating current (AC).Direct current (DC). Indicates hazardous v

Página 57 - Chapter 4 Service 55

Testing theMultiplexerYou can use the tests and checks in Chapter 2, "Verification Tests," and theinformation in Table 4-4 to isolate the pr

Página 58 - Troubleshooting Techniques

Repair and Maintenance GuidelinesThis section provides guidelines for handling the relay multiplexersincluding:• ESD precautions• Soldering printed ci

Página 59 - (E1355A/56A)

Component Locators and Schematic DiagramsThe following tables list Component Locator Diagrams and SchematicDiagrams provided with this manual. Refer t

Página 60 - Testing the

Table 4-7. Terminal Module Component Locator and Schematic Diagram Drawings (for all E1343A/44A/45A/47A/55A/56A Relay Multiplexers, regardless of seri

Página 61 - Post-Repair Safety

Notes62 Service Chapter 4

Página 62 - 60 Service Chapter 4

Appendix AVerification Tests - C ProgramsFunctional Verification TestThese programs are designed to do the Functional Verification Tests foundin Chapt

Página 63 - Chapter 4 Service 61

PerformanceVerificationTestsThese programs are designed to do the Performance Verification Testsfound in Chapter 2, "Verification Tests."Exa

Página 64 - 62 Service Chapter 4

iprintf (dm, "TRIG SGL\n"); iscanf (dm, "%lf", &R[i]); iscanf (dm, "%t", cr); printf ("\nR

Página 65 - Appendix A

j = 100; for (i = 0;i <= 2; i++) { iprintf (id, "CLOS (@%u)\n", j); iprintf (dm, "TRIG SGL\n"); iscanf

Página 66 - Verification

getchar (); iprintf (id, "CLOS (@%u)\n", j); iprintf (dm, "TRIG SGL\n"); iscanf (dm, "%lf", &value1[0]

Página 67

Declaration of Conformity Declarations of Conformity for this product and for other Agilent products may be downloaded from the Internet. There are

Página 68

Example: DCIsolation TestThis example performs DC Isolation Tests for HI to Chassis, HI to LO, andHI to GU (Guard). printf ("\n\nChannels 90-9

Página 69

iprintf (dm, "TRIG SGL\n"); iscanf (dm, "%t", reading); printf ("\nDC Isolation -- HI to CHASSIS "); print

Página 70 - Isolation Test

Notes70 Example C Programs Appendix A

Página 71

Appendix BBackdating InformationIntroductionThis chapter contains information for ordering replaceable parts for theE1343A/44A/45A/47A/55A/56A Relay

Página 72

E1345-80001. To order only the Terminal Card, use E1345-66510. To orderonly the Terminal Case Assembly, use E1300-84401.Table B-1. Relay Multiplexer A

Página 73 - Backdating Information

ReferenceDesignatorPartNumberQty Part DescriptionJ1J2J3K0-K15K90-K931252-15961252-37121251-60010490-170721120COMPONENT ASSY (CONT’D...HP E1345-66201)C

Página 74

ReferenceDesignatorPartNumberQty Part DescriptionA1U1-U3A1U5A1U6A1U7-U8A1U91858-00691820-40571820-67311820-30791820-308131121COMPONENT ASSY (CONT’D...

Página 75

Table B-3. E1343A, E1344A Replaceable Parts(See applicable Serial Number ranges listed on the first page of this chapter))ReferenceDesignatorPart Numb

Página 76

ReferenceDesignatorPart Number Qty. Part DescriptionCOMPONENT ASSY (CONT’D...E1343-66201)A1A1C1A1C2-C3A1C5A1C6A1C7-C9A1C11A1C17A1C38-C42A1C44A1C45-C46

Página 77

ReferenceDesignatorPart Number Qty. Part DescriptionCOMPONENT ASSY (CONT’D...E1343-66201)A1U6A1U7-U8A1U9A1U10-U11A1U12A1U151820-67311820-30791820-3081

Página 78

Notes6 E1343A/44A/45A/47A/55A/56A Service Manual

Página 79

Parts LocatorsFigures B-1 through B-3 show locations of replaceable parts for theE1343A, E1344A, E1345A, E1347A, E1355A, and E1356A RelayMultiplexers

Página 80

Figure B-2. E1343A/44A/45A/47A Terminal Module Mechanical PartsAppendix B Backdating Information 79

Página 81

Figure B-3. E1355A/56A Terminal Module Mechanical Parts80 Backdating Information Appendix B

Página 83

Copyright © Agilent Technologies, Inc. 2005 *E1345-90013* E1345-90013

Comentários a estes Manuais

Sem comentários