Module 5
Basic Measurement
5-35
Simplified Diagram
x1
Buffer
SMU
Guard
Sense
DUT
Force
V
Rs
10K
Guard Connection for Kelvin connection
With the Kelvin connection, sensing is done at the DUT, eliminating the fraction of an ohm of cable
resistance. The internal sensing resistor Rs is the only feedback path without the Kelvin connection.
Note that the B1500A operates just fine without the sense cable. This is important to know because
in general you do not need the sensing Kelvin connection. Most MOS measurements are high
impedance and the residual cable loss is insignificant.