Dual Electrical Channel Modules 86112A 54754A
Electrical Channel Bandwidth 12.4 and 20 GHz 12.4 and 18 GHz
Transition Time (10% to 90%, 28.2 ps (12.4 GHz); 28.2 ps (12.4 GHz);
calculated from TR = 0.35/BW) 17.5 ps (20 GHz) 19.4 ps (18 GHz)
RMS Noise
Characteristic 0.25 mV (12.4 GHz); 0.25 mV (12.4 GHz);
0.5 mV (20 GHz) 0.5 mV (18 GHz)
Maximum 0.5 mv (12.4 GHz); 0.5 mv (12.4 GHz);
1 mV (20 GHz) 1 mV (18 GHz)
Scale Factor
Minimum 1 mV/division
Maximum 100 mV/division
DC Accuracy (single marker) ±0.4% of full scale ±0.4% of full scale or marker reading
±2mV ±1.5% of (reading-channel offset), 12.4 GHz (whichever is greater)
±0.4% of full scale ±2 mV ±1.2% of (reading-channel offset)
±2 mV ±3% of (reading-channel offset), 20 GHz
CW Offset Range (referenced from
center of screen) ±500 mV ±500 mV
Input Dynamic Range (relative to
channel offset) ±400 mV ±400 mV
Maximum Input Signal ±2 V (+16 dBm) ±2 V (+16 dBm)
Nominal Impedance 50 ohm 50 ohm
Reflections (for 30 ps rise time) 5% 5%
Electrical Input 3.5 mm (male) 3.5 mm (male)
Module Specifications: Dual Electrical
Dual Electrical Channel Modules 86117A 86117B 86118A
Electrical Channel Bandwidth 30 and 50 GHz 30, 50 amd 65 GHz 50 and 70 GHz
Transition Time (10% to 90%, 11.7 ps (30 GHz) 11.7 ps (30 GHz)
calculated from TR = 0.35/BW) 7 ps (50 GHz) 7 ps (50 GHz)
5.4 ps (65 GHz)
RMS Noise
Characteristic 0.4 mV (30 GHz) 0.4 mV (30 GHz) 0.7 mV (50 GHz)
0.6 mV (50 GHz) 0.5 mV (50 GHz) 1.3 mV (70 GHz)
1.3 mV (65 GHz)
Maximum 0.7 mv (30 GHz); 0.7 mv (30 GHz) 1.8 mV (50 GHz)
1.0 mV (50 GHz 0.9 mV (50 GHz) 2.5 mV (70 GHz)
2.2 mV (65 GHz)
Scale Factor
Minimum 1 mV/division
Maximum 100 mV/division
DC Accuracy (single marker) ±0.4% of full scale ±0.4% of full scale ±0.4% of full scale
±2 mV ±1.2% of (reading- ±2 mV ±1.2% of (reading- ±1.5 mV ±1% of (reading-
channel offset) (30 GHz) channel offset) (30 GHz) channel offset) (50 GHz)
±0.4% of full scale ±0.4% of full scale ±0.4% of full scale
±2 mV ±2% of (reading- ±2 mV ±2% of (reading- ±1.5 mV ±3% of (reading-
channel offset) (50 GHz) channel offset) (50 GHz) channel offset) (70 GHz)
±0.4% of full scale
±2 mV ±2% of (reading-
channel offset) (65 GHz)
CW Offset Range (referenced from
center of screen) ±500 mV
Input Dynamic Range (relative to
channel offset) ±400 mV
Maximum Input Signal ±2 V (+16 dBm)
Nominal Impedance 50 ohm
Reflections (for 30 ps rise time) 5%
Electrical Input 2.4 mm (male) 1.85 mm (male) 1.85 mm (male)