Agilent Technologies 81134A Manual de Serviço Página 52

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 51
SCPI Commands Reference SOURce Commands
52 Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007
[:SOURce]:FREQuency[:CW|:FIXed]
Syntax
[:SOUR]:FREQ[:CW|:FIX][?]
Form Set & Query
Description This command programs the internal clock frequency, and also
selects the internal clock as time base if it is not already selected.
Parameter Numeric [GHz|MHz|kHz|Hz] | MIN|MAX
Value coupling Period = 1 / Frequency
*RST value 15.0E6 Hz
Specified limits 15E6 ... 3.35E9 Hz, with overclocking up to 3.35E9 Hz
Example Select the clock with frequency 1.2 GHz.
:FREQ 1.2GHz
Vista de página 51
1 2 ... 47 48 49 50 51 52 53 54 55 56 57 ... 103 104

Comentários a estes Manuais

Sem comentários