From Trigger / Start / X Marker Field
The from Trigger/Start/X marker field is used to access the selection pop-up
for the start point of the X and O marker occurrence counters.
The start points available for the X-marker are either the trace start point or
the trigger point. The start points available for the O-marker are the trace
start point, trigger point, or the X-marker.
If the marker pattern can not be found, a message appears at the top of the
display indicating the search failed. If the O-marker is referenced from the X
marker, and the X-marker is not found, the search for both markers will fail.
Trigger / Start / X-Marker Field
First occurrence
of X pattern after
Trigger / Start / X Marker field
The Listing Menu
From Trigger / Start / X Marker Field
13–8