54 4000 X-Series Oscilloscopes Advanced Training Guide
3 Advanced Triggering, Search & Navigation, and Segmented Acquisition Labs
Your scope’s display should now look similar to Figure 31. Using the
scope’s default rising edge type trigger, we can see two distinctly different
rising edge speeds on this waveform. Also notice that the slower
transitioning edge appears dimmer. This is because the slower
transitioning edges occur less often than the faster transitioning edges.
Perhaps these slower transitioning edges are in violation of meeting
minimum required specifications. Let’s first set up the scope to measure
the rise times of this signal, and then we will set up the scope to uniquely
trigger on the violation edges.
9 Press [Meas]; then tap the Ty pe softkey.
10 Select the Rise Time measurement; Once on, tap Statistics and then
Display On.
Figure 31 Scope display reveals two different rising edge speeds when triggering on any
rising edge.