Agilent Technologies 22D Guia do Utilizador Página 251

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 276
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 250
7-13
Agilent 54640-series Performance
Characteristics
* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and
±10 ° C from firmware calibration temperature.
Acquisition: Analog Channels
Max Sample rate 2 GSa/s interleaved, 1 GSa/s each channel
Max Memory Depth 8 MB interleaved, 4 MB each channel
Vertical Resolution 8 bits
Peak Detection 1 ns @ max sample rate
Averages selectable from 2, 4, 8, 16, 32, 64 ...to 16383
High Resolution Mode 12 bits of resolution when
100 us/div, average mode with average = 1
Filter: Sinx/x interpolation (single shot BW = sample rate/4 or bandwidth of scope, whichever
is less) with vectors on.
Acquisition: Digital Channels (on 54641D and 54642D only)
Max Sample Rate 1 GSa/s
Max Memory Depth 4 MB
Vertical Resolution 1 bit
Glitch Detection
(min pulse width)
5 ns
Vista de página 250
1 2 ... 246 247 248 249 250 251 252 253 254 255 256 ... 275 276

Comentários a estes Manuais

Sem comentários