Module 7
Measurement Functions
7-48
SMU1 Function to VAR1
Vh = 0 to 2.5 V
Bias Hold to ON
Output Value to STOP
Bias Hold Exercise
Channel and Measurement Setup
Set the SMU1 Function to VAR1. The VAR1 output 0 to 2.5 V is enough to light the LED.
Set the Bias Hold after Measurement field to ON to enable the bias hold function. And set the
Output Value after Measurement field to STOP. This setup keeps 2.5 V output between sweep