Module 7
Measurement Functions
7-22
-90 ms =< Hold Time =< -0.1 ms, 0.1 ms step
Bias starts at 0 s
Negative Hold Time
Available for Interval < 2 ms
For the sampling measurement, the available hold time value is 0 to 655.35 s, in 10 ms step.
However, the negative hold time -90 ms to -0.1 ms, in 0.1 ms step is also available for the high speed
sampling which the Interval is less than 2 ms.
This measurement result example is obtained by the setup shown in the following pages.