114 Chapter 6
Channel Analyzer Measurements
Making Adjacent Channel Power (ACP) Measurements
Channel Analyzer Measurements
Press Meas Setup, Limits, Power Limits, Center Chan High Limit, 10, dBm,
Center Chan Low Limit, 25, dBm, Adj Chan 1 High Limit, 35, dB, and
Adj Chan 2 High Limit, 65, dB.
In Figure 6-4 notice that ACP 2 Low and ACP 2 High have both failed,
however all other channels have passed.
Figure 6-4 Setting Offset Limits