Agilent Technologies N5416A Manual de Serviço Página 67

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 236
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 66
Device Hi-Speed Tests 3
Notes on USB Electrical Compliance Testing 67
Connecting the Equipment
The USB automated test application will prompt you to perform these
connection steps:
1 Attach the 5V power supply to the E2649- 66403 (E2645- 66503 if you
are using the old fixture) Device Receiver test fixture (J5).
a Verify the green Power LED is lit.
b Leave the TEST switch at the OFF position (S1).
c The yellow LED should be off.
2 Connect the [INIT PORT] of the fixture to a Hi- Speed port on the Test
Bed Computer, using the 5 meter USB cable.
3 Connect the [TEST PORT] of the fixture to the device under test, using
the 4" USB cable.
4 Connect the Agilent 113xA differential probe to the test fixture at TP2,
using the 01131- 68703 header adapter (the damped header adapter is
only needed if you are using the old test fixture).
Vista de página 66
1 2 ... 62 63 64 65 66 67 68 69 70 71 72 ... 235 236

Comentários a estes Manuais

Sem comentários