Device Hi-Speed Tests 3
Notes on USB Electrical Compliance Testing 55
2 Click OK to close the Test Instructions dialog.
EL_27 Device CHIRP Response to Reset from Hi-Speed Operation
The USB automated test application will prompt you to perform these
steps:
1 On the Device Test Menu of the HS Electrical Test Tool, select RESET
from the Device Command drop down menu. Click [EXECUTE] once to
reset the device operating in high speed. The captured transition should
be as in the figure below.